MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s10096-011-1356-1

Медицинская библиотека · European journal of clinical microbiology & infectious diseases : official publication of the European Society of Clinical Microbiology · 2011

Comparative evaluation of Vitek 2 identification and susceptibility testing of Gram-negative rods directly and isolated from BacT/ALERT-positive blood culture bottles.

Munoz-Dávila M J, Yagüe G, Albert M, García-Lucas T

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

The performance of Vitek 2 was evaluated for the identification and susceptibility testing of Gram-negative bacilli directly from positive blood cultures bottles. Direct inoculation of the positive blood cultures with the Vitek cards ID-GN and AST-NO58 was compared with the standard inoculation method based on the sub-culture of the positive blood culture to agar. A total of 142 blood cultures were included in the study; of those, 119 were from patients' clinical samples, while 23 were artificially prepared with strains showing different mechanisms of resistance. A total of 136 (95.8%) strains were correctly identified to the species level, only 2 (1.4%) were mis-identified and 4 (2.8%) were not identified. Susceptibility results were available for all isolates tested against 17 antibiotics, thus, resulting in a total of 2,414 isolate/anti-microbial combinations. The error rate was 2.8% (67/2,414) overall; 0.6% (14/2,414) very major errors, 0.1% (3/2,414) major errors and 2.1% (50/2,414) minor errors. The direct method detected 88.5% (22/25) of the strains producing extended-spectrum beta-lactamases (ESBLs). The susceptibility agreement among the added strains with ESBL, AMPc hyperproduction, resistance to ceftazidime, carbapenems and cefepime was very high. Direct identification and susceptibility testing gave rapid and reliable results, reducing by 24 h the turnaround time of the microbiology laboratory.

Для цитирования: Munoz-Dávila M J, Yagüe G, Albert M, García-Lucas T. Comparative evaluation of Vitek 2 identification and susceptibility testing of Gram-negative rods directly and isolated from BacT/ALERT-positive blood culture bottles.. European journal of clinical microbiology & infectious diseases : official publication of the European Society of Clinical Microbiology. 2011. DOI: 10.1007/s10096-011-1356-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.