Study on the effect of three-low technique in aortic computed tomography angiography for patients with body mass index <30 kg/m2: a randomized controlled prospective trial.

Chen Zhijian, Wang Shaogang, Zhang Ran, Chen Ying, Deng Xiaoyi, Zhang Xueqin, Zhang Tao

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Aortic computed tomography (CT) angiography (ACTA) is a key noninvasive tool for assessing aortic diseases, which typically requires high tube voltage and large volumes of contrast agent, thus potentially increasing radiation exposure and the risk of contrast agent-induced nephropathy. This study aimed to explore the effect of three-low technique (low tube voltage, low contrast agent injection rate, and low contrast agent volume) in CTA for patients with a body mass index (BMI) <30 kg/m2, thus this technique may be used as an alternative to conventional ACTA.

METHODS

Eighty patients with a BMI less than 30 kg/m2 undergoing CTA examination with a Philips iCT 256-slice scanner were randomly assigned into two groups according to the examination technique used: three-low group (n=40): tube voltage: 100 kV, contrast agent injection rate: 3 mL/s, contrast agent volume: 50 mL; control group (n=40): tube voltage: 120 kV, injection rate: 5 mL/s, contrast agent: 80 mL. Aortic CT value and image noise were detected, the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were computed, and the subjective score of the image quality was evaluated. The effective radiation dose was computed according to a formula.

RESULTS

Compared with the control group, the effective radiation dose was decreased by 76.2% (10.32±0.46 vs. 2.46±0.71 mSv), while the aortic CT value was increased by approximately 13% (360.14±56.52 vs. 402.12±88.34 HU), the image noise (22.23±5.34 vs. 13.32±3.54) was higher, while the contrast agent volume (80 vs. 50 mL), SNR (24.2±7.4 vs. 20.6±6.1) and CNR (22.4±6.5 vs. 18.3±6.2) were lower in the three-low group, and all these differences between the two groups were statistically significant (all P<0.05). There was no statistically significant difference in the subjective score of image quality (4.4±0.8 vs. 4.2±0.9) between the two groups (P>0.05).

CONCLUSIONS

Three-low technique has a good application value in CTA for patients with a BMI of lower than 30 kg/m2; the effective radiation dose was decreased by 76.2%, the contrast agent volume was decreased by 37.5%, and all images met the diagnostic requirements.

Для цитирования: Chen Zhijian, Wang Shaogang, Zhang Ran, Chen Ying, Deng Xiaoyi, Zhang Xueqin, Zhang Tao. Study on the effect of three-low technique in aortic computed tomography angiography for patients with body mass index <30 kg/m2: a randomized controlled prospective trial.. Quantitative imaging in medicine and surgery. 2025. DOI: 10.21037/qims-2025-203.