Comparison of liver MRI R2(FerriScan®) VS liver MRI T2* as a measure of body iron load in a cohort of beta thalassaemia major patients.

Padeniya Padmapani, Siriwardana Shirom, Ediriweera Dileepa, Samarasinghe Nayana, Silva Sasanka, Silva Ishari, Ahamed Nizri, Niriella Madunil, Premawardhena Anuja

Письмо в редакциюПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

To compare the similarity of the non-patented T2* and the high cost patented R2 (Ferriscan®) MRI techniques in the measurement of liver iron concentration (LIC) in heavily transfused patients with thalassaemia major in a real- life Sri Lankan hospital setup. We compared LIC measured by MRI, obtained 2 weeks apart, using both T2* and R2 techniques in 15 patients with beta thalassaemia major. They all had a history of > 100 units of blood transfusions life long and also a history of sub optimal chelation. MRI R2 and MRI T2* scan values showed a negative correlation (co-rrelation coefficient = - 0.63, p = 0.01) This correlation was strong in lower LICs and progressively decreased with upper LIC values. Thus a significant discrepancy was observed between median values of two MRI technologies (p = 0.0005) with T2* tending to underestimate iron overload especially in those with very high LIC identified by R2. The lack of concordance of T2* and R2 especially in those with very high reading on R2 suggest the potential errors in interpretations that can occur in "non-expert centres"; which are likely to lead to errors in clinical judgement on the intensity of chelation therapy needed.

Для цитирования: Padeniya Padmapani, Siriwardana Shirom, Ediriweera Dileepa, Samarasinghe Nayana, Silva Sasanka, Silva Ishari, Ahamed Nizri, Niriella Madunil, Premawardhena Anuja. Comparison of liver MRI R2(FerriScan®) VS liver MRI T2* as a measure of body iron load in a cohort of beta thalassaemia major patients.. Orphanet journal of rare diseases. 2020. DOI: 10.1186/s13023-020-1301-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.