MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2016.01.007

Медицинская библиотека · JACC. Cardiovascular interventions · 2016

The Characteristics of In-Stent Restenosis After Drug-Eluting Stent Implantation in Femoropopliteal Lesions and 1-Year Prognosis After Repeat Endovascular Therapy for These Lesions.

Iida Osamu, Takahara Mitsuyoshi, Soga Yoshimitsu, Hirano Keisuke, Yamauchi Yasutaka, Zen Kan, Kawasaki Daizo, Nanto Shinsuke, Yokoi Hiroyoshi, Uematsu Masaaki, ZEPHYR Investigators

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study sought to investigate the characteristics of in-stent restenosis (ISR) after drug-eluting stent (DES) implantation for femoropopliteal (FP) lesions, and to examine 1-year prognosis after repeat endovascular therapy (re-EVT) for these DES-ISR.

BACKGROUND

The morphology of DES-ISR and its association with clinical outcomes after re-EVT have not been well examined.

METHODS

This was a subanalysis of the ZEPHYR (Zilver PTX for the femoral artery and proximal popliteal artery) study. The current study included 210 cases with loss of patency confirmed 1 year after DES implantation. Morphology of DES-ISR was classified into the following subgroups: class I, focal lesions (≤50 mm in length), class II, diffuse lesions (>50 mm in length), and class III, totally occluded ISR. One-year prognosis after re-EVT for DES-ISR was assessed by restenosis and major adverse limb events (MALE).

RESULTS

Classes I, II, and III accounted for 50%, 25%, and 25% of DES-ISR, respectively. Factors associated with the morphology of DES-ISR were the presence of chronic total occlusion and the size of the external elastic membrane area before DES implantation (p = 0.009 and 0.017). Compared with the class I restenotic lesion, the class II and III lesions had a significantly higher risk of restenosis (74% and 78% vs. 53%: p = 0.048 and 0.019, respectively) and MALE (56% and 56% versus 32%: p = 0.025 and 0.022, respectively) 1 year after re-EVT.

CONCLUSIONS

We evaluated the characteristics of ISR after DES implantation for FP lesions and 1-year prognosis of re-EVT for DES-ISR. The morphology of DES-ISR had a significant association with 1-year prognosis after re-EVT.

Для цитирования: Iida Osamu, Takahara Mitsuyoshi, Soga Yoshimitsu, Hirano Keisuke, Yamauchi Yasutaka, Zen Kan, Kawasaki Daizo, Nanto Shinsuke, Yokoi Hiroyoshi, Uematsu Masaaki, ZEPHYR Investigators. The Characteristics of In-Stent Restenosis After Drug-Eluting Stent Implantation in Femoropopliteal Lesions and 1-Year Prognosis After Repeat Endovascular Therapy for These Lesions.. JACC. Cardiovascular interventions. 2016. DOI: 10.1016/j.jcin.2016.01.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.