DOI 10.1016/j.hrthm.2020.08.009
Primary prevention implantable cardioverter-defibrillators in hypertrophic cardiomyopathy-Are there predictors of appropriate therapy?
Аннотация
BACKGROUND
Identifying patients with hypertrophic cardiomyopathy (HCM) who warrant a primary prevention implantable cardioverter-defibrillator (ICD) is crucial. ICDs are effective in terminating life-threatening arrhythmias; however, ICDs carry risks of complications.
OBJECTIVE
The purpose of this study was to assess the incidence and predictors of appropriate ICD therapies, inappropriate shocks, and device-related complications in patients with HCM and primary prevention ICDs.
METHODS
All patients with HCM who underwent primary prevention ICD implantation at Toronto General Hospital between September 2000 and December 2017 were identified. Therapies (shocks or antitachycardia pacing) for ventricular tachycardia >180 beats/min or ventricular fibrillation were considered appropriate.
RESULTS
Three hundred two patients were followed for a mean 6.1 ± 4.3 years (1801 patient-years of follow-up). Thirty-eight patients (12.6%) received at least 1 appropriate ICD therapy (2.3%/y); the 5-year cumulative probability of receiving appropriate ICD therapy was 9.6%. None of the conventional risk factors nor the European Society of Cardiology risk score was associated with appropriate ICD therapy. In multivariable analysis, age < 40 years at implantation and atrial fibrillation were independent predictors of appropriate ICD therapy. In a subgroup of patients who had undergone cardiac magnetic resonance imaging before ICD implantation, severe late gadolinium enhancement was the strongest predictor of appropriate ICD therapies. ICD-related complications or inappropriate shocks occurred in 87 patients (28.8%), with an inappropriate shock rate of 2.1%/y; the 5-year cumulative probability was 10.7%.
CONCLUSION
The incidence of appropriate ICD therapies in patients with HCM and primary prevention ICDs is lower than previously reported; a high proportion of patients suffer from an ICD-related complication. Traditional risk factors have low predictive utility. Severe late gadolinium enhancement, atrial fibrillation, and young age are important predictors of ventricular tachyarrhythmias in HCM.