MedElevator Войти в исследования
← Поиск публикаций

DOI 10.21037/jtd-24-853

Медицинская библиотека · Journal of thoracic disease · 2025

Application of novel dark-quencher labeled probes in multiplex qRT-PCR assays for rapid detection of SARS-CoV-2 variants.

Zeng Zhiqi, Yang Jie, Dai Jun, Zou Lirong, Lin Zhengshi, Liu Yong, Guan Wenda, Li Feng, Zheng Kui, Yuan Shuai, Sun Fangfang, He Fengxia, Hong Ye, Li Hui, Liu Wei, Men Guangqi, Zhang Xinyue, Lan Yun, Deng Xizi, Li Liya, Lin Yaqing, Lai Honghao, Qian Peng, Fan Qinghong, Jiang Mengling, Li Jiaojiao, Tang Guofang, Mo Qiaohui, Deng Xiaoyan, Huang Jicheng, Deng Xiaoling, Yang Zifeng

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Coronavirus disease 2019 (COVID-19) is an acute infectious disease caused by the new coronavirus, the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Because SARS-CoV-2 frequently mutates, it creates a number of variants that must be distinguished and tracked using a rapid detection technique. At present, the identification of virus variants often requires sequencing of the viral genome with sophisticated techniques which are costly and time-consuming. On the other hand, the quantitative reverse transcription-polymerase chain reaction (qRT-PCR) method used to diagnose SARS-CoV-2 infection has been widely applied worldwide amid COVID-19 pandemic. Due to the lower specificity and sensitivity in detecting different strains using multiple qRT-PCR, we aimed to develop novel dark quencher (DQ) labeled probes to improve the performance of multiple qRT-PCR. DQ probes are dihydropyrroloindole carboxylate (DPI3)-analogue.

METHODS

We first tested their amplification efficiency and specificity, on detecting single nucleotide polymorphism through qRT-PCR, and the simultaneous detection efficiency of multiple SARS-CoV-2 mutation sites. The DQ labeled probes were further applied in multiplex qRT-PCR assays, and the method was validated on SARS-CoV-2 positive clinical samples for its sensitivity and specificity.

RESULTS

DQ probes exhibited better specificity and sensitivity than the TaqMan® Minor Groove Binder (MGB) and TaqMan probes. Great analytical sensitivity (limit of detection of 250 copies/mL), good specificity (no cross-reaction with other pathogens), and great clinical performance (99.4-100% consistency with next-generation sequencing) were demonstrated by the designed multiplex qRT-PCR tests.

CONCLUSIONS

Our novel DQ-probe/multiplex qRT-PCR assay provides a rapid and simple method to quickly distinguish SARS-CoV-2 variants, we were able to quickly identify SARS-CoV-2 variants (Delta and Omicron BA.1, BA.1.1, BA.2, BA.2.12.1, BA.3, BA.4, and BA.5) that target nine specific mutation sites in the ORF, N, NSP1, NSP3, and S genes.

Для цитирования: Zeng Zhiqi, Yang Jie, Dai Jun, Zou Lirong, Lin Zhengshi, Liu Yong, Guan Wenda, Li Feng, Zheng Kui, Yuan Shuai, Sun Fangfang, He Fengxia, Hong Ye, Li Hui, Liu Wei, Men Guangqi, Zhang Xinyue, Lan Yun, Deng Xizi, Li Liya, Lin Yaqing, Lai Honghao, Qian Peng, Fan Qinghong, Jiang Mengling, Li Jiaojiao, Tang Guofang, Mo Qiaohui, Deng Xiaoyan, Huang Jicheng, Deng Xiaoling, Yang Zifeng. Application of novel dark-quencher labeled probes in multiplex qRT-PCR assays for rapid detection of SARS-CoV-2 variants.. Journal of thoracic disease. 2025. DOI: 10.21037/jtd-24-853.