MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3389/fcvm.2022.892122

Медицинская библиотека · Frontiers in cardiovascular medicine · 2022

Initial Experience in Transvenous Implantation of a Left Ventricular Lead With a Novel Venogram Balloon Catheter.

Duan Jiangbo, Yang Dandan, He Jinshan, Li Xuebin, Wang Long, Wu Cuncao, Li Ding, Ze Feng, Yuan Cuizhen, Zhou Jingliang, Zhou Xu

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

AIM: The most challenging and time-consuming stage of cardiac resynchronization therapy (CRT) device implantation is coronary sinus (CS) cannulation and left ventricular epicardial electrode implantation. This paper reports the initial clinical experience of CS cannulation and left ventricular lead implantation guided by a novel venogram balloon catheter (Lee's venogram balloon catheter).

METHODS AND

RESULTS

Consecutive patients eligible for CRT were deemed suitable for this novel venogram balloon catheter. Parameters such as left ventricular lead implantation time, procedure time, and fluoroscopy time were recorded. CS cannulation with LV lead implantation guided by Lee's venogram balloon catheter was successful in all 5 patients, including 4 challenging cases. The total fluoroscopy and procedural durations were 5.0 ± 3.0 and 57.4 ± 12.5 min, respectively. No adverse catheter-related events occurred during the procedures.

CONCLUSION

This initial study of an innovative venogram balloon catheter demonstrated that it greatly facilitated CS cannulation and successful LV lead placement in all patients undergoing CRT system implantation. This significantly shortened the learning curve and showed a decrease in left ventricular lead implantation time, procedure time, and fluoroscopy time.

Для цитирования: Duan Jiangbo, Yang Dandan, He Jinshan, Li Xuebin, Wang Long, Wu Cuncao, Li Ding, Ze Feng, Yuan Cuizhen, Zhou Jingliang, Zhou Xu. Initial Experience in Transvenous Implantation of a Left Ventricular Lead With a Novel Venogram Balloon Catheter.. Frontiers in cardiovascular medicine. 2022. DOI: 10.3389/fcvm.2022.892122.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.