MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/jce.15724

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2022

Esophageal temperature management during cryoballoon ablation for atrial fibrillation.

Sink Joshua, Nimmagadda Kiran, Zhao Manyun, Andrei Adin-Cristian, Gay Hawkins, Kaplan Rachel M, Gao Xu, Pfenniger Anna, Patil Kaustubha D, Arora Rishi, Kim Susan S, Chicos Alexandru B, Lin Albert C, Passman Rod S, Knight Bradley P, Verma Nishant

Рандомизированное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

Esophageal thermal injury (ETI) is a well-recognized complication of atrial fibrillation (AF) ablation. Previous studies have demonstrated that direct esophageal cooling reduces ETI during radiofrequency AF ablation. The purpose of this study was to evaluate the use of an esophageal warming device to prevent ETI during cryoballoon ablation (CBA) for AF.

METHODS

This prospective, double-blinded study enrolled 42 patients with symptomatic AF undergoing CBA. Patients were randomized to the treatment group with esophageal warming (42°C) using recirculated water through a multilumen, silicone tube inserted into the esophagus (EnsoETM®; Attune Medical) (WRM) or the control group with a luminal single-electrode esophageal temperature monitoring probe (LET). Patients underwent upper endoscopy esophagogastroduodenoscopy (EGD) the following day. ETI was classified into four grades.

RESULTS

Baseline patient characteristics were similar between groups. Procedural characteristics including number of freezes, total freeze time, early freeze terminations, coldest balloon temperature, procedure duration, posterior wall ablation, and proton pump inhibitor and transesophageal echocardiogram use before procedure were not different between groups. The EGD was completed in 40/42 patients. There was significantly more ETI in the WRM group compared to the LET group (n = 8 [38%] vs. n = 1 [5%], p = 0.02). All ETI lesions were grade 1 (erythema) or 2 (superficial ulceration). Total freeze time in the left inferior pulmonary vein was predictive of ETI (360 vs. 300 s, p = 0.03).

CONCLUSION

Use of a luminal heat exchange tube for esophageal warming during CBA for AF was paradoxically associated with a higher risk of ETI.

Для цитирования: Sink Joshua, Nimmagadda Kiran, Zhao Manyun, Andrei Adin-Cristian, Gay Hawkins, Kaplan Rachel M, Gao Xu, Pfenniger Anna, Patil Kaustubha D, Arora Rishi, Kim Susan S, Chicos Alexandru B, Lin Albert C, Passman Rod S, Knight Bradley P, Verma Nishant. Esophageal temperature management during cryoballoon ablation for atrial fibrillation.. Journal of cardiovascular electrophysiology. 2022. DOI: 10.1111/jce.15724.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.