DOI 10.1016/j.prosdent.2025.06.010
Optimized design of a tooth-simulating screw for enhanced accuracy in robot-assisted implant placement in completely edentulous jaws.
Аннотация
STATEMENT OF PROBLEM: The absence of natural teeth in computer-assisted complete arch implant surgery presents challenges as natural teeth as anatomic landmarks are lacking.
PURPOSE: The purpose of this in vitro study was to evaluate implant placement accuracy using a robotic-assisted system with a novel tooth-simulating screw simulating natural teeth and optimizing their distribution in a completely edentulous jaw model.
MATERIAL AND
METHODS
A total of 12 identical 3-dimensionally (3D) printed models were used, with 4 implants placed in each. The models were divided into 4 groups based on the number (3 or 4) and distribution (dispersed or localized) of tooth-simulating screws. Cone beam computed tomography (CBCT) data of the tooth-simulating screw in a mandible model were imported into a robotic planning software program for registration, implants were placed under full robotic guidance, and deviations between planned and actual positions were measured by superimposing pre-operative and postoperative CBCT data. All measurements were recorded as absolute values and presented as mean ±standard deviation. Two-way ANOVA was used for data analyses.
RESULTS
The use of the tooth-simulating screw to assist registration and tracker fixation resulted in high accuracy. No significant differences were observed among the groups in terms of platform and apex deviation (P>.05); however, angular deviation was affected (P<.05). Tooth- simulating screws in a dispersed distribution demonstrated a significantly lower angular deviation compared with localized groups. Among the dispersed groups, 3 tooth-simulating screws (0.50 ±0.19 degree) exhibited higher accuracy than 4 tooth-simulating screws (0.82 ±0.26 degree).
CONCLUSIONS
The novel tooth-simulating screw used in robot-assisted implant surgery achieved high accuracy. Dispersed tooth-simulating screw placement led to higher accuracy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.