Knee kinematics are not different between asymmetrical and symmetrical tibial baseplates in total knee arthroplasty: A fluoroscopic analysis of step-up and lunge motions.

Koster Lennard A, Kaptein Bart L, van der Linden-van der Zwaag Enrike H M J, Nelissen Rob G H H

Рандомизированное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: This clinical fluoroscopy study investigated knee kinematics of two different cemented fixed-bearing, posterior-stabilised (PS) total knee arthroplasty (TKA) designs: an asymmetric tibial component including an asymmetric insert designed to optimise personalised balance and fit and its precursor symmetrical design with symmetric insert.

METHODS

A consecutive series of patients (16 TKAs from each treatment group) participating in a randomised controlled trial comparing TKA migration was included. The exclusion criterion was the use of walking aids. Flat-panel fluoroscopic recordings of step-up and lunge motions were acquired 1-year postoperatively. Medial and lateral contact points (CPs) were determined to calculate CP displacement, femoral axial rotation and pivot position. Using linear mixed-effects modelling techniques, kinematics between TKA designs were compared.

RESULTS

During knee extension between 20° flexion and full extension, the CPs moved anteriorly combined with a small internal femoral rotation (a screw-home mechanism). Whereas CP movement was reversed: femoral rollback, external femoral rotation while flexing the knee between full extension and 20° knee flexion, At larger flexion angles, femoral axial rotation (FAR) occurred around a lateral pivot point both during step-up and lunge. The symmetric design had a 2.3° larger range of FAR compared to the asymmetric design during lunge (p = 0.02). All other kinematics were comparable.

CONCLUSION

Despite the differences in design, this study showed that the asymmetric and symmetric PS TKA designs had mostly comparable knee kinematics during step-up and lunge motions. It is therefore expected that the functionality of the successor TKA design is similar to that of its precursor design.

LEVEL OF EVIDENCE: Level II, prospective comparative study.

Для цитирования: Koster Lennard A, Kaptein Bart L, van der Linden-van der Zwaag Enrike H M J, Nelissen Rob G H H. Knee kinematics are not different between asymmetrical and symmetrical tibial baseplates in total knee arthroplasty: A fluoroscopic analysis of step-up and lunge motions.. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA. 2024. DOI: 10.1002/ksa.12125.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.