Comparison of tracheal versus esophageal temperatures during laparoscopic surgery.
Наблюдательное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Lee Eun Kyung, Bang Yu Jeong, Kim Jeayoun, Ahn Hyun Joo
Аннотация
PURPOSE: Recently, endotracheal tubes with an embedded temperature sensor in the inner surface of the tube cuff (temperature tracheal tubes) have been developed. We sought to assess whether temperature tracheal tubes show a good agreement with esophageal temperature probes during surgery.
METHODS: We enrolled 40 patients who underwent laparoscopic surgery in an observational study. The tracheas of all patients were intubated with a temperature tracheal tube, and an esophageal temperature probe was inserted into the esophagus. Tracheal and esophageal temperatures were recorded at 15-min intervals until the end of surgery. Temperatures from both devices were analyzed using Bland-Altman analysis, four-quadrant plots, and polar plots.
RESULTS: We analyzed 261 data points from 36 patients. Temperatures ranges were 34.2 °C to 36.6 °C for the tracheal temperature tube and 34.7 °C to 37.2 °C for the esophageal temperature probe. Bland-Altman analysis showed an acceptable agreement between the two devices, with an overall mean bias (95% limit of agreement) of -0.3 °C (-0.8 °C to 0.1 °C) and a percentage error of 3%; the trending ability (temperature changes over time) between the two devices showed a concordance rate of 94% in four-quadrant plot (cut-off ≥ 92%), but this was higher than the acceptable mean angular bias of 177° (cut-off < ± 5°) and radial limits of agreement of 52° (cut-off < ± 30°) in the polar plot. Bronchoscopy during extubation and patient interviews at six hours postoperatively revealed no serious injuries related to the use of the temperature tracheal tube.
CONCLUSION: The temperature tracheal tube showed an acceptable overall mean bias of -0.3 °C and a percentage error of 3%, but incompatible trending ability with the esophageal temperature probe.
STUDY REGISTRATION: cris.nih.go.kr (KCT0007265); 22 April 2022.
METHODS: We enrolled 40 patients who underwent laparoscopic surgery in an observational study. The tracheas of all patients were intubated with a temperature tracheal tube, and an esophageal temperature probe was inserted into the esophagus. Tracheal and esophageal temperatures were recorded at 15-min intervals until the end of surgery. Temperatures from both devices were analyzed using Bland-Altman analysis, four-quadrant plots, and polar plots.
RESULTS: We analyzed 261 data points from 36 patients. Temperatures ranges were 34.2 °C to 36.6 °C for the tracheal temperature tube and 34.7 °C to 37.2 °C for the esophageal temperature probe. Bland-Altman analysis showed an acceptable agreement between the two devices, with an overall mean bias (95% limit of agreement) of -0.3 °C (-0.8 °C to 0.1 °C) and a percentage error of 3%; the trending ability (temperature changes over time) between the two devices showed a concordance rate of 94% in four-quadrant plot (cut-off ≥ 92%), but this was higher than the acceptable mean angular bias of 177° (cut-off < ± 5°) and radial limits of agreement of 52° (cut-off < ± 30°) in the polar plot. Bronchoscopy during extubation and patient interviews at six hours postoperatively revealed no serious injuries related to the use of the temperature tracheal tube.
CONCLUSION: The temperature tracheal tube showed an acceptable overall mean bias of -0.3 °C and a percentage error of 3%, but incompatible trending ability with the esophageal temperature probe.
STUDY REGISTRATION: cris.nih.go.kr (KCT0007265); 22 April 2022.