MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/jcp.2003.014910

Медицинская библиотека · Journal of clinical pathology · 2004

Use of tissue microarray for interlaboratory validation of HER2 immunocytochemical and FISH testing.

Kay E, O'Grady A, Morgan J M, Wozniak S, Jasani B

Сравнительное исследованиеОценочное исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

AIMS: To evaluate the use of tissue microarray (TMA) technology as a validation tool for HER2 testing by both immunocytochemistry (ICC) and fluorescence in situ hybridisation (FISH) in the diagnostic setting.

METHODS

TMA constructs from 57 cases of breast cancer were evaluated for HER2 (by ICC and FISH) by two centres. The results were compared.

RESULTS

There was a high level of concordance for both ICC and FISH. In five "discrepant" cases only three would have had a potential impact on patient management.

CONCLUSIONS

Validation of HER2 analysis in the clinical setting by ICC and FISH is essential. The use of TMAs provides for an economy of scale and would be practical in the setting of interlaboratory and intralaboratory validation. It is suggested that routine HER2 ICC and FISH should continue to be performed in laboratories on whole sections. Following this, TMAs would be constructed for all cases of breast cancer. ICC and FISH would be performed on these to validate the results. The TMAs would be available for circulation to other centres for validation purposes. The standardisation of testing between centres, the potential difficulty of minimum case numbers, and the workload issues surrounding validation would all be facilitated by this approach.

Для цитирования: Kay E, O'Grady A, Morgan J M, Wozniak S, Jasani B. Use of tissue microarray for interlaboratory validation of HER2 immunocytochemical and FISH testing.. Journal of clinical pathology. 2004. DOI: 10.1136/jcp.2003.014910.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.