BACKGROUND

The 308-nm excimer laser and 308-nm excimer lamp have both been shown to be effective in treating vitiligo but a direct comparison has never been performed.

OBJECTIVES

To test the equivalence of these two devices for treating nonsegmental vitiligo.

PATIENTS AND

METHODS

A randomized monocentric study was undertaken. One lesion was treated with the 308-nm excimer laser and its counterpart with the 308-nm excimer lamp. Lesions were treated twice weekly with the same dose on both sides for a total of 24 sessions. The evaluation was done by two independent physicians blinded to the treatment on direct light and ultraviolet light photos.

RESULTS

Twenty patients were included: 17 completed the study and 104 lesions were treated. The two treatments showed similar results in terms of efficacy for a repigmentation of at least 50% (P = 0.006). The lamp induced more erythema than the laser.

CONCLUSIONS

The 308-nm excimer lamp and laser showed a similar efficacy in treating vitiligo. For the same fluence, the lamp induced more erythema suggesting photobiological differences between the two devices.

Для цитирования: Le Duff F, Fontas E, Giacchero D, Sillard L, Lacour J-P, Ortonne J-P, Passeron T. 308-nm excimer lamp vs. 308-nm excimer laser for treating vitiligo: a randomized study.. The British journal of dermatology. 2010. DOI: 10.1111/j.1365-2133.2010.09778.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.