Impact of the repositionable Evolut R CoreValve system on the need for a permanent pacemaker after transcatheter aortic valve implantation in patients with severe aortic stenosis.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ojeda Soledad, Hidalgo Francisco, Romero Miguel, Mazuelos Francisco, Suárez de Lezo Javier, Martín Ernesto, Lostalo Adrián, Luque Aurora, González Rafael, Fernández Ana, López-Aguilera José, Segura José, Guerrero Noelia, Pan Manuel
Аннотация
OBJECTIVES: To compare the incidence of permanent pacemaker implantation (PPI) with the CoreValve and Evolut R prostheses, to evaluate the implantation depth with both types of prostheses, and to study factors predicting the need for PPI.
BACKGROUND: The Evolut R CoreValve can be recaptured and repositioned during deployment, allowing a more precise implantation.
METHODS: A total of 208 patients treated with CoreValve and 137 patients treated with Evolut R were analyzed. The depth of the prosthesis in the LVOT was measured by angiography in the annular perpendicular view projection after deploymen in all patients.
RESULTS: Baseline conduction abnormalities were comparable between the groups (85/208, 40.9% vs. 53/137, 38.7%; p = 0.69). The mean prosthesis depth was 10.3 ± 8.6 mm in the CoreValve group and 5.5 ± 2.7 mm in the Evolut R group; p < 0.0001. Conduction disturbances after valve implantation were more frequent with the CoreValve (new-onset left bundle branch block: 93, 44.7% vs. 16, 11.7%; p < 0.05, first-degree atrioventricular block: 23, 11.1% vs. 5, 3.6%; p < 0.05). In addition, the incidence of PPI was significantly lower with Evolut R (45, 21.6% vs. 15, 10.9%; p = 0.01). The predictors of the need for PPI were the mean depth of the prosthesis (OR: 1.13, 95% CI: 1.06-1.21; p < 0.0001) and prior right bundle branch block (OR 10.22, 95% CI: 4.62-22.63; p < 0.0001).
CONCLUSIONS: The recapturable capability of the Evolut R system allowed for higher and precise valve implantation. This fact had an impact on the reduction in the need for PPI.
BACKGROUND: The Evolut R CoreValve can be recaptured and repositioned during deployment, allowing a more precise implantation.
METHODS: A total of 208 patients treated with CoreValve and 137 patients treated with Evolut R were analyzed. The depth of the prosthesis in the LVOT was measured by angiography in the annular perpendicular view projection after deploymen in all patients.
RESULTS: Baseline conduction abnormalities were comparable between the groups (85/208, 40.9% vs. 53/137, 38.7%; p = 0.69). The mean prosthesis depth was 10.3 ± 8.6 mm in the CoreValve group and 5.5 ± 2.7 mm in the Evolut R group; p < 0.0001. Conduction disturbances after valve implantation were more frequent with the CoreValve (new-onset left bundle branch block: 93, 44.7% vs. 16, 11.7%; p < 0.05, first-degree atrioventricular block: 23, 11.1% vs. 5, 3.6%; p < 0.05). In addition, the incidence of PPI was significantly lower with Evolut R (45, 21.6% vs. 15, 10.9%; p = 0.01). The predictors of the need for PPI were the mean depth of the prosthesis (OR: 1.13, 95% CI: 1.06-1.21; p < 0.0001) and prior right bundle branch block (OR 10.22, 95% CI: 4.62-22.63; p < 0.0001).
CONCLUSIONS: The recapturable capability of the Evolut R system allowed for higher and precise valve implantation. This fact had an impact on the reduction in the need for PPI.
Предметные рубрики MeSH
AgedAged, 80 and overAortic ValveAortic Valve StenosisArrhythmias, CardiacCardiac Pacing, ArtificialFemaleHeart Valve ProsthesisHumansMalePacemaker, ArtificialProsthesis DesignRetrospective StudiesRisk FactorsSeverity of Illness IndexTime FactorsTranscatheter Aortic Valve ReplacementTreatment Outcome