MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3389/fneur.2026.1865673

Медицинская библиотека · Frontiers in neurology · 2026

Dual-target VIM-PSA deep brain stimulation using octopolar electrodes: surgical technique and feasibility study.

Triguero-Cueva Lucía, Martínez-Barbero José Pablo, Madrid-Navarro Carlos Javier, Pérez-Navarro María José, Mínguez-Castellanos Adolfo, Marín-Romero Bartolomé, Iáñez-Velasco Benjamín, Escamilla-Sevilla Francisco, Katati Majed Jouma

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

BACKGROUND/

OBJECTIVES

Deep brain stimulation (DBS) is an established surgical treatment for essential tremor (ET). Both the thalamic ventral intermediate nucleus (VIM) and the posterior subthalamic area (PSA) are effective tremor targets can be approached through a single trajectory. This technical feasibility report describes a single-trajectory dual-target VIM-PSA DBS surgical protocol using octopolar electrodes and evaluates stereotactic accuracy, anatomical contact distribution, and preliminary surgical safety.

METHODS

Eleven patients with medication-refractory ET underwent bilateral DBS implantation within the framework of an ongoing randomized, double-blind, crossover clinical trial comparing VIM-DBS and PSA-DBS. Preoperative planning was performed using Brainlab Elements, and postoperative reconstruction was performed using Guide™ XT after fusion of preoperative MRI with one-month postoperative CT. Clinical and electrical data (efficacy outcomes, stimulation thresholds, side-effect thresholds, active contact selection, energy consumption, habituation, patient-reported outcomes, and programming burden) were not assessed in this report.

RESULTS

A total of 22 octopolar DBS leads were implanted. Postoperative reconstruction confirmed anatomical contact distribution across the reconstructed VIM model and the PSA/cZI target region. Mean planned-to-implanted deviation across contacts was 1.28 mm for the left electrodes and 1.50 mm for the right electrodes. One patient developed transient symptomatic perielectrode edema; no intracerebral hemorrhage or any other major surgery-related adverse event was observed during early postoperative follow-up.

CONCLUSION

Single-trajectory dual-target VIM-PSA DBS using octopolar electrodes was successfully performed with acceptable stereotactic accuracy and without major surgery-related adverse events in this small cohort. The approach may increase anatomical coverage and postoperative programming options. Clinical data remain to be determined.

Для цитирования: Triguero-Cueva Lucía, Martínez-Barbero José Pablo, Madrid-Navarro Carlos Javier, Pérez-Navarro María José, Mínguez-Castellanos Adolfo, Marín-Romero Bartolomé, Iáñez-Velasco Benjamín, Escamilla-Sevilla Francisco, Katati Majed Jouma. Dual-target VIM-PSA deep brain stimulation using octopolar electrodes: surgical technique and feasibility study.. Frontiers in neurology. 2026. DOI: 10.3389/fneur.2026.1865673.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.