DOI 10.1111/j.1600-051x.2005.00863.x

Медицинская библиотека · Journal of clinical periodontology · 2006

In vitro calculus detection with a moved smart ultrasonic device.

Meissner Grit, Oehme Bernd, Strackeljan Jens, Kocher Thomas

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

STUDY

OBJECTIVE

The objective of subgingival instrumentation of periodontally diseased root surfaces is to remove the adhering microbial biofilm and calcified deposits. Recently, we have described an automated calculus detection system under static conditions. Clinically however, the tip of the system has to be moved over tooth surfaces. It was thus necessary to study the entire system in motion.

METHODS

The detection device is based on a conventional dental piezoelectric ultrasonic handpiece with a conventional scaler insert. The impulse response of the mechanical oscillation system is analysed by a fuzzy logic-based computerized algorithm, which classifies various surfaces. The present study investigates dental surface recognition properties of the new system with the tip being moved over teeth surfaces in vitro. Following a training set of 7977 measurements (3960 calculus, 4017 cement) on 200 extracted teeth, 1363 measurements were conducted on 34 teeth unknown to the system.

RESULTS

The surfaces cementum and calculus were correctly classified in 78% within the training set and in 81% within the set unknown, with a kappa value of 0.68.

CONCLUSION

It was shown that this method of automatic recognition of tooth surfaces is able to distinguish between different tooth surfaces in vitro independently from tip movements.

Для цитирования: Meissner Grit, Oehme Bernd, Strackeljan Jens, Kocher Thomas. In vitro calculus detection with a moved smart ultrasonic device.. Journal of clinical periodontology. 2006. DOI: 10.1111/j.1600-051x.2005.00863.x.