Transcatheter Mitral Valve Replacement With the Transseptal EVOQUE System.

Webb John, Hensey Mark, Fam Neil, Rodés-Cabau Josep, Daniels David, Smith Robert, Szeto Wilson, Boone Robert, Ye Jian, Moss Robert, Peterson Mark, Ong Geraldine, Paradis Jean-Michel, Dagenais François, Wintzer-Wehekind Jérôme, Shafi Bilal, Nair Deepu, Szerlip Molly, Grayburn Paul, Silvestry Frank, Koulogiannis Konstantinos, Hermiller James, Guyton Robert, Herrmann Howard

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

The aim of this study was to examine the initial experience with a novel transseptal transcatheter mitral valve replacement (TMVR) system.

BACKGROUND

Transseptal TMVR may offer a less invasive option than surgery for mitral regurgitation (MR) with greater efficacy and fewer anatomic limitations than transcatheter repair.

METHODS

Patients were treated with the EVOQUE TMVR system from September 2018 to October 2019. Key inclusion criteria were moderate or greater MR, New York Heart Association functional class ≥II, and high or prohibitive surgical risk. The primary outcome was technical success, defined by Mitral Valve Academic Research Consortium criteria.

RESULTS

Fourteen patients were treated, all with at least moderate to severe MR. The median age was 84 years, and the median Society of Thoracic Surgeons score was 4.6%. MR was degenerative in 4 (28.6%), functional in 3 (21.4%), and mixed in 7 (50%). Technical success was achieved in 13 patients (92.9%), and 1 patient was converted to surgery. At 30 days there was 1 noncardiovascular mortality (7.1%), 2 strokes (14.3%), no myocardial infarctions, and no rehospitalizations. Two patients (14.3%) underwent paravalvular leak closure. One patient (7.1%) underwent alcohol septal ablation for left ventricular outflow tract obstruction. Including the 2 patients with paravalvular leak closure, MR was mild or less in all implanted patients at 30 days, with no MR in 10 (83.3%). Mean mitral gradient was 5.8 mm Hg (median). New York Heart Association functional class improved to ≤II in 9 patients (81.8%).

CONCLUSIONS

This first-in-human experience has demonstrated the feasibility of the transseptal EVOQUE TMVR system. Further clinical studies are required to establish safety and clinical outcomes.

Для цитирования: Webb John, Hensey Mark, Fam Neil, Rodés-Cabau Josep, Daniels David, Smith Robert, Szeto Wilson, Boone Robert, Ye Jian, Moss Robert, Peterson Mark, Ong Geraldine, Paradis Jean-Michel, Dagenais François, Wintzer-Wehekind Jérôme, Shafi Bilal, Nair Deepu, Szerlip Molly, Grayburn Paul, Silvestry Frank, Koulogiannis Konstantinos, Hermiller James, Guyton Robert, Herrmann Howard. Transcatheter Mitral Valve Replacement With the Transseptal EVOQUE System.. JACC. Cardiovascular interventions. 2020. DOI: 10.1016/j.jcin.2020.06.040.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.