MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/ejcts/ezaa256

Медицинская библиотека · European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery · 2021

Safety and performance of a novel transventricular beating heart mitral valve repair system: 1-year outcomes.

Gammie James S, Bartus Krzysztof, Gackowski Andrzej, Szymanski Piotr, Bilewska Agata, Kusmierczyk Mariusz, Kapelak Boguslaw, Rzucidlo-Resil Jolanta, Duncan Alison, Yadav Rashmi, Livesey Steve, Diprose Paul, Gerosa Gino, D'Onofrio Augusto, Pittarello Demetrio, Denti Paolo, La Canna Giovanni, De Bonis Michele, Alfieri Ottavio, Hung Judy, Kolsut Piotr, D'Ambra Michael N

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The objective of this study was to evaluate the safety and performance of a novel, beating heart procedure that enables echocardiographic-guided beating heart implantation of expanded polytetrafluoroethylene (ePTFE) artificial cords on the posterior mitral leaflet of patients with degenerative mitral regurgitation.

METHODS

Two prospective multicentre studies enrolled 13 (first-in-human) and 52 subjects, respectively. Patients were treated with the HARPOON beating heart mitral valve repair system. The primary (30-day) end point was successful implantation of cord(s) with mitral regurgitation reduction to ≤moderate. An independent core laboratory analysed echocardiograms.

RESULTS

Of 65 patients enrolled, 62 (95%) achieved technical success, 2 patients required conversion to open surgery and 1 procedure was terminated. The primary end point was met in 59/65 (91%) patients. Among the 62 treated patients, the mean procedural time was 2.1 ± 0.5 h. Through discharge, there were no deaths, strokes or renal failure events. At 1 year, 2 of the 62 patients died (3%) and 8 (13%) others required reoperations. At 1 year, 98% of the patients with HARPOON cords were in New York Heart Association class I or II, and mitral regurgitation was none/trace in 52% (n = 27), mild in 23% (n = 12), moderate in 23% (n = 12) and severe in 2% (n = 1). Favourable cardiac remodelling outcomes at 1 year included decreased end-diastolic left ventricular volume (153 ± 41 to 119 ± 28 ml) and diameter (53 ± 5 to 47 ± 6 mm), and the mean transmitral gradient was 1.4 ± 0.7 mmHg.

CONCLUSIONS

This initial clinical experience with the HARPOON beating heart mitral valve repair system demonstrates encouraging early safety and performance.

CLINICAL REGISTRATION NUMBERS: NCT02432196 and NCT02768870.

Для цитирования: Gammie James S, Bartus Krzysztof, Gackowski Andrzej, Szymanski Piotr, Bilewska Agata, Kusmierczyk Mariusz, Kapelak Boguslaw, Rzucidlo-Resil Jolanta, Duncan Alison, Yadav Rashmi, Livesey Steve, Diprose Paul, Gerosa Gino, D'Onofrio Augusto, Pittarello Demetrio, Denti Paolo, La Canna Giovanni, De Bonis Michele, Alfieri Ottavio, Hung Judy, Kolsut Piotr, D'Ambra Michael N. Safety and performance of a novel transventricular beating heart mitral valve repair system: 1-year outcomes.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2021. DOI: 10.1093/ejcts/ezaa256.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.