DOI 10.1016/j.jclinane.2024.111403
Thoracic impedance pneumography in propofol-sedated patients undergoing percutaneous endoscopic gastrostomy (PEG) placement in gastrointestinal endoscopy: A prospective, randomized trial.
Аннотация
STUDY
OBJECTIVE
To assess the efficacy of an ECG-based method called thoracic impedance pneumography to reduce hypoxic events in endoscopy.
DESIGN
This was a single center, 1:1 randomized controlled trial.
SETTING
The trial was conducted during the placement of percutaneous endoscopic gastrostomy (PEG).
PATIENTS: 173 patients who underwent PEG placement were enrolled in the present trial. Indication was oncological in most patients (89%). 58% of patients were ASA class II and 42% of patients ASA class III.
INTERVENTIONS: Patients were randomized in the standard monitoring group (SM) with pulse oximetry and automatic blood pressure measurement or in the intervention group with additional thoracic impedance pneumography (TIM). Sedation was performed with propofol by gastroenterologists or trained nurses.
MEASUREMENTS: Hypoxic episodes defined as SpO2 < 90% for >15 s were the primary endpoint. Secondary endpoints were minimal SpO2, apnea >10s/>30s and incurred costs.
MAIN
RESULTS
Additional use of thoracic impedance pneumography reduced hypoxic episodes (TIM: 31% vs SM: 49%; p = 0.016; OR 0.47; NNT 5.6) and elevated minimal SpO2 per procedure (TIM: 90.0% ± 8.9; SM: 84.0% ± 17.6; p = 0.007) significantly. Apnea events >10s and > 30s were significantly more often detected in TIM (43%; 7%) compared to SM (1%; 0%; p < 0.001; p = 0.014) resulting in a time advantage of 17 s before the occurrence of hypoxic events. As a result, adjustments of oxygen flow were significantly more often necessary in SM than in TIM (p = 0.034) and assisted ventilation was less often needed in TIM (2%) compared with SM (9%; p = 0.053). Calculated costs for the additional use of thoracic impedance pneumography were 0.13$ (0.12 €/0.11 £) per procedure.
CONCLUSIONS
Additional thoracic impedance pneumography reduced the quantity and extent of hypoxic events with less need of assisted ventilation. Supplemental costs per procedure were negligible.
KEY WORDS: thoracic impedance pneumography, capnography, sedation, monitoring, gastrointestinal endoscopy, percutaneous endoscopic gastrostomy.