DOI 10.1093/bjr/tqaf193

Медицинская библиотека · The British journal of radiology · 2025

An assessment of the average glandular dose of breast cone-beam CT in China based on clinical patients.

Xue Ke, Xu Hui, Yue Bao-Rong, Ding Yan-Qiu, Ye Zhao-Xiang

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To quantify the non-compressed breast characteristics of females and the typical scanning parameters of breast cone-beam computed tomography (CBCT), and to provide a preliminary assessment of the average glandular dose (AGD) of breast CBCT in China.

METHODS

This study retrospectively analysed 3D breast imaging data from 1590 Chinese women who underwent non-enhanced breast CBCT scans in different regions between October 2020 and November 2023. Initially, a semiautomatic measurement tool integrated into the breast CBCT system was used to quantify the breast characteristics for each patient, including effective diameter at the chest-wall (Deff), chest-wall to nipple length (CNL), breast volume (BV), glandular volume (GV), and volumetric breast density (VBD). Then, the AGD for each patient was calculated using the dose conversion factor, combining breast characteristics and scanning parameters.

RESULTS

The overall median values of Deff, CNL, BV, GV, and VBD reflecting breast characteristics were 12.6 cm, 6.8 cm, 370.6 cm3, 46.7 cm3, and 14.4%, respectively. The tube voltage was fixed at 49 kVp, and the typical tube current was 50 mA. Under the typical scanning parameters, the AGD ranged from 3.9 to 5.1 mGy (median 4.9 mGy).

CONCLUSIONS

The typical AGD for non-enhanced breast CBCT scanning is ∼1.81× higher than the cumulative radiation dose from 2 digital mammography views in China.

ADVANCES IN KNOWLEDGE: This study assessed the AGD of patients undergoing breast CBCT examination based on a large amount of clinical data, providing valuable guidance to clinicians in choosing appropriate breast imaging modalities.

Для цитирования: Xue Ke, Xu Hui, Yue Bao-Rong, Ding Yan-Qiu, Ye Zhao-Xiang. An assessment of the average glandular dose of breast cone-beam CT in China based on clinical patients.. The British journal of radiology. 2025. DOI: 10.1093/bjr/tqaf193.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.