The Journal of antimicrobial chemotherapy · 2021

Development and validation of a lateral flow immunoassay for rapid detection of VanA-producing enterococci.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Oueslati Saoussen, Volland Hervé, Cattoir Vincent, Bernabeu Sandrine, Girlich Delphine, Dulac Ducan, Plaisance Marc, Laroche Maxime, Dortet Laurent, Simon Stéphanie, Naas Thierry

Аннотация

BACKGROUND: VRE are nosocomial pathogens with an increasing incidence in recent decades. Rapid detection is crucial to reduce their spread and prevent infections and outbreaks.

OBJECTIVES: To evaluate a lateral flow immunoassay (LFIA) (called NG-Test VanA) for the rapid and reliable detection of VanA-producing VRE (VanA-VRE) from colonies and broth.

METHODS: NG-Test VanA was validated on 135 well-characterized enterococcal isolates grown on Mueller-Hinton (MH) agar (including 40 VanA-VRE). Different agar plates and culture broths widely used in routine laboratories for culture of enterococci were tested.

RESULTS: All 40 VanA-VRE clinical isolates were correctly detected in less than 15 min irrespective of the species expressing the VanA ligase and the medium used for bacterial growth. No cross-reaction was observed with any other clinically relevant ligases (VanB, C1, C2, D, E, G, L, M and N). Overall, the sensitivity and specificity of the assay were 100% for VanA-VRE grown on MH agar plates. NG-Test VanA accurately detects VanA-VRE irrespective of the culture medium (agar and broth). Band intensity was increased when using bacteria grown on vancomycin-containing culture media or on MH close to the vancomycin disc as a consequence of VanA induction. The limit of detection of the assay was 6.3 × 106 cfu per test with bacteria grown on MH plates and 4.9 × 105 cfu per test with bacteria grown on ChromID® VRE plates.

CONCLUSIONS: NG-Test VanA is efficient, rapid and easy to implement in the routine workflow of a clinical microbiology laboratory for the confirmation of VanA-VRE.