Active middle ear implants-Enhanced efficiency of oval window stimulation with larger piston size.

Ghoncheh Mohammad, Schurzig Daniel, Stauske David, Lenarz Thomas, Prenzler Nils, Liu Houguang, Maier Hannes

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

INTRODUCTION

Direct piston-stimulation of the oval window with a middle ear implant, following a stapedotomy or stapedectomy, is one treatment option for patients with otosclerosis. Here, we experimentally investigated whether increasing the surface area of the actuated piston enhances transmission efficiency in oval window (OW) stimulation.

METHODS

Fresh-frozen human cadaveric temporal bones (N = 14) were used to evaluate the output of different oval window couplers (OWCs) of A = 0.38 mm², 1.26 mm2, 3.06 mm2 by intracochlear pressure difference measurements between scala vestibuli (PSV) and scala tympani (PST). Following stapedectomy, a piezoelectric transducer was used for OWC actuation. Additionally, a numerical model was also developed to estimate the output of the floating mass transducer (FMT, MEDEL, Austria).

RESULTS

Across frequencies, the average pressure difference was 18 and 8 dB higher for the large and medium OWC compared to the small OWC, respectively. Statistical analyses showed significant differences (p < 0.05) between piston diameters across most frequencies. Numerical simulation results using the volume displacement at the OW showed smaller output differences from 0.3 to 10 kHz, indicating that leakage at open gaps after stapedectomy has to be taken into account.

CONCLUSION

Our findings confirm that, for active stimulation with an actuator, increasing the surface area of the OWC enhances output efficiency proportionally. Nevertheless, volume leakage at open gaps to the OW may contribute additionally.

Для цитирования: Ghoncheh Mohammad, Schurzig Daniel, Stauske David, Lenarz Thomas, Prenzler Nils, Liu Houguang, Maier Hannes. Active middle ear implants-Enhanced efficiency of oval window stimulation with larger piston size.. Hearing research. 2025. DOI: 10.1016/j.heares.2025.109411.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.