JACC. Heart failure · 2015

Echocardiographic Ramp test for continuous-flow left ventricular assist devices: do loading conditions matter?

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Adatya Sirtaz, Holley Christopher T, Roy Samit S, Yarmohammadi Hirad, Feng Amy, Eckman Peter, Colvin-Adams Monica, John Ranjit, Masri Carolina

Аннотация

OBJECTIVES: This study investigated whether continuous AI and/or elevated mean arterial pressure (MAP) were associated with false positive results for flow obstruction in echocardiographic ramp speed tests in patients with a continuous-flow left ventricular assist device.

BACKGROUND: Failure to reduce the left ventricular end-diastolic diameter (LVEDD) with increasing device speeds in a ramp test is predictive of pump obstruction. Aortic insufficiency (AI) or increased MAP can diminish the ability to unload the left ventricle.

METHODS: LVEDD was plotted against device speed, and a linear function slope was calculated. A flat LVEDD slope (≥-0.16) was considered abnormal (suggestive of obstruction). Ramp test results were compared in patients with or without either AI or increased MAP at baseline speed, and receiver-operator characteristic (ROC) curves were constructed for predictors of device obstruction. Device thrombosis was confirmed by direct visualization of clot at explantation or on inspection by the manufacturer.

RESULTS: Of 78 ramp tests (55 patients), 36 were abnormal (18 true positive, 18 false positive), and 42 were normal (37 true negative, 5 false negative). In patients with AI, LVEDD slope was -0.14 ± 0.17, which was consistent with device obstruction (vs. -0.25 ± 0.11 in patients without AI; p < 0.001), despite no difference in mean lactate dehydrogenase concentration between the 2 groups (1,301 ± 1,651 U/l vs. 1,354 ± 1,365 U/l; p = 0.91). Area under the ROC curve (AUC) for LVEDD slope was 0.76 and improved to 0.88 after removal of patients with AI from the study. LVEDD slope in patients with MAP ≥85 mm Hg was similar to that for device obstruction (-0.18 ± 0.07) and was abnormal in 6 of the 12 ramp tests performed. Combining LVEDD slope with lactate dehydrogenase concentration increased the AUC to 0.96 as an indicator of device obstruction.

CONCLUSIONS: Abnormal loading conditions due to AI or elevated MAP may result in false positive ramp tests.