m+Medelevator.RU
← Поиск публикаций
Nuklearmedizin. Nuclear medicine · 2007

Equipment-independent reference values for dopamine transporter imaging with 123I-FP-CIT.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Koch W, Hornung J, Hamann C, Pöpperl G, Tatsch K

Аннотация

AIM: Reliable reference values are helpful to interpret and compare the results of dopamine transporter imaging with SPECT. Since semi-quantitative reference values cannot be easily transferred between imaging equipments, this study aimed to establish equipment independent normal values for the true striatal binding of 123I-FP-CIT.

PATIENTS, METHODS: Specific striatal FP-CIT binding of 6 healthy volunteers and 26 patients with essential tremor were used to generate a reference range by applying an equipment specific resolution dependent factor to compensate for recovery effects. This factor has been determined previously by a series of standardized phantom measurements of an anthropomorphic basal ganglia phantom. Herewith, the resulting DAT binding values represent the expected true specific binding in the striatum.

RESULTS: On average, true specific striatal binding was 5.83 +/- 0.96 in healthy controls, 5.25 +/- 0.67 in patients with essential tremor and 5.36 +/- 0.75 in the entire study cohort.

CONCLUSION: These preliminary results may serve as a basis for the generation of a generally accepted equipment independent reference range for dopamine transporter imaging with 123I-FP-CIT. By a simple phantom measurement that can be accomplished within one day factors related to specific imaging equipment and processing can be corrected for, resulting in specific binding values which may enable a more standardized interpretation of dopamine transporter scans.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты