The Journal of thoracic and cardiovascular surgery · 2014
Transapical sutureless aortic valve implantation under magnetic resonance imaging guidance: Acute and short-term results.
Оценочное исследованиеСтатьяResearch Support, N.I.H., IntramuralНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Horvath Keith A, Mazilu Dumitru, Cai Junfeng, Kindzelski Bogdan, Li Ming
Аннотация
OBJECTIVES: Despite the increasing success and applicability of transcatheter aortic valve replacement, 2 critical issues remain: the durability of the valves, and the ideal imaging to aid implantation. This study was designed to investigate the transapical implantation of a device of known durability using real-time magnetic resonance imaging (MRI) guidance.
METHODS: A sutureless aortic valve was used that employs a self-expanding nitinol stent and is amenable to transapical delivery. MRI (1.5-T) was used to identify the anatomic landmarks in 60-kg Yucatan swine. Prostheses were loaded into an MRI-compatible delivery device with an active guidewire to enhance visualization. A series of acute feasibility experiments were conducted (n = 10). Additional animals (n = 6) were allowed to survive and had follow-up MRI scans and echocardiography at 90 days postoperatively. Postmortem gross examination was performed.
RESULTS: The valve was MRI compatible and created no significant MRI artifacts. The 3 commissural struts were visible on short-axis view; therefore, coronary ostia obstruction was easily avoided. The average implantation time was 65 seconds. Final results demonstrated stability of the implants with preservation of myocardial perfusion and function over 90 days: the ejection fraction was 48% ± 15%; the peak gradient was 17.3 ± 11.3 mm Hg; the mean gradient was 9.8 ± 7.2 mm Hg. Mild aortic regurgitation was seen in 4 cases, trace in 1 case, and a severe central jet in 1 case. Prosthesis positioning was evaluated during gross examination.
CONCLUSIONS: We demonstrated that a sutureless aortic valve can be safely and expeditiously implanted through a transapical approach under real-time MRI guidance. Postimplantation results showed a well-functioning prosthesis, with minimal regurgitation, and stability over time.
METHODS: A sutureless aortic valve was used that employs a self-expanding nitinol stent and is amenable to transapical delivery. MRI (1.5-T) was used to identify the anatomic landmarks in 60-kg Yucatan swine. Prostheses were loaded into an MRI-compatible delivery device with an active guidewire to enhance visualization. A series of acute feasibility experiments were conducted (n = 10). Additional animals (n = 6) were allowed to survive and had follow-up MRI scans and echocardiography at 90 days postoperatively. Postmortem gross examination was performed.
RESULTS: The valve was MRI compatible and created no significant MRI artifacts. The 3 commissural struts were visible on short-axis view; therefore, coronary ostia obstruction was easily avoided. The average implantation time was 65 seconds. Final results demonstrated stability of the implants with preservation of myocardial perfusion and function over 90 days: the ejection fraction was 48% ± 15%; the peak gradient was 17.3 ± 11.3 mm Hg; the mean gradient was 9.8 ± 7.2 mm Hg. Mild aortic regurgitation was seen in 4 cases, trace in 1 case, and a severe central jet in 1 case. Prosthesis positioning was evaluated during gross examination.
CONCLUSIONS: We demonstrated that a sutureless aortic valve can be safely and expeditiously implanted through a transapical approach under real-time MRI guidance. Postimplantation results showed a well-functioning prosthesis, with minimal regurgitation, and stability over time.
Предметные рубрики MeSH
Anatomic LandmarksAnimalsAortic ValveAortic Valve InsufficiencyArtifactsCoronary CirculationFeasibility StudiesHeart Valve ProsthesisHeart Valve Prosthesis ImplantationMagnetic Resonance Imaging, CineMagnetic Resonance Imaging, InterventionalModels, AnimalOperative TimePredictive Value of TestsProsthesis DesignStroke VolumeSus scrofaTime Factors