The Thoracic and cardiovascular surgeon · 2020
Oversized versus Non-oversized Prosthesis: Midterm Outcomes after Transcatheter Aortic Valve Replacement Using SAPIEN 3 Valve.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Useini Dritan, Beluli Blerta, Christ Hildegard, Mügge Andreas, Patsalis Polykarpos, Schlömicher Markus, Haldenwang Peter, Bechtel Matthias, Strauch Justus
Аннотация
BACKGROUND: A certain degree of prosthesis oversizing is recommended for the SAPIEN 3 (S3; Edwards Lifesciences Corp., Irvine, California, United States) to ensure device success. We aimed to investigate midterm outcomes in patients who received oversized (OS) S3 valve after transapical-transcatheter aortic valve replacement (TA-TAVR).
METHODS: Out of 122 patients with aortic stenosis who underwent TA-TAVR using S3 at our institution, 42 received OS prosthesis. We used computed tomography (CT) derived effective diameter to assess oversizing. We defined oversizing if the labeled diameter of the selected valve for implantation was ≥2 mm bigger than the effective annulus diameter calculated by the annulus area. We conducted a midterm follow-up and compared the OS cohort with the non-OS (nOS) cohort.
RESULTS: The study groups showed similar risk score and age (STS [Society of Thoracic Surgery] score: 5.4 ± 3; mean age: 80.7 ± 5.7). The 30-day mortality was 7.1% in OS versus 2.4% in nOS. The 30-day all-stroke was 2.4% in OS versus 0% in nOS. The 1- and 3-year all-cause mortality were 28.5 and 42.8% in OS versus 21.9 and 26.8% in nOS, respectively. Midterm freedom from death and from cardiocerebral events was similar in both groups. Moderate/severe paravalvular leakage occurred in 0% in OS versus 5.4% in nOS. The postdilation rate was 7.1% in OS versus 15.3% in nOS. The rate of new permanent pacemaker implantation (PPI) was 15.7% in OS versus 9.3% in nOS. The STS score was detected as an independent predictor of mortality.
CONCLUSION: Oversizing reduces the risk of device failure and intraprocedural postdilation but increases the risk of PPI. Early and midterm morbidity and mortality after OS and nOS with S3 are comparable.
METHODS: Out of 122 patients with aortic stenosis who underwent TA-TAVR using S3 at our institution, 42 received OS prosthesis. We used computed tomography (CT) derived effective diameter to assess oversizing. We defined oversizing if the labeled diameter of the selected valve for implantation was ≥2 mm bigger than the effective annulus diameter calculated by the annulus area. We conducted a midterm follow-up and compared the OS cohort with the non-OS (nOS) cohort.
RESULTS: The study groups showed similar risk score and age (STS [Society of Thoracic Surgery] score: 5.4 ± 3; mean age: 80.7 ± 5.7). The 30-day mortality was 7.1% in OS versus 2.4% in nOS. The 30-day all-stroke was 2.4% in OS versus 0% in nOS. The 1- and 3-year all-cause mortality were 28.5 and 42.8% in OS versus 21.9 and 26.8% in nOS, respectively. Midterm freedom from death and from cardiocerebral events was similar in both groups. Moderate/severe paravalvular leakage occurred in 0% in OS versus 5.4% in nOS. The postdilation rate was 7.1% in OS versus 15.3% in nOS. The rate of new permanent pacemaker implantation (PPI) was 15.7% in OS versus 9.3% in nOS. The STS score was detected as an independent predictor of mortality.
CONCLUSION: Oversizing reduces the risk of device failure and intraprocedural postdilation but increases the risk of PPI. Early and midterm morbidity and mortality after OS and nOS with S3 are comparable.
Предметные рубрики MeSH
AgedAged, 80 and overAortic ValveAortic Valve StenosisFemaleHeart Valve ProsthesisHumansMalePostoperative ComplicationsProsthesis DesignProsthesis FailureRecovery of FunctionRetrospective StudiesRisk AssessmentRisk FactorsTime FactorsTomography, X-Ray ComputedTranscatheter Aortic Valve ReplacementTreatment Outcome