Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis: the detect supraventricular tachycardia study.

Friedman Paul A, McClelland Robyn L, Bamlet William R, Acosta Helbert, Kessler David, Munger Thomas M, Kavesh Neal G, Wood Mark, Daoud Emile, Massumi Ali, Schuger Claudio, Shorofsky Stephen, Wilkoff Bruce, Glikson Michael

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Delivery of inappropriate shocks caused by misdetection of supraventricular tachycardia (SVT) remains a substantial complication of implanted cardioverter/defibrillator (ICD) therapy. Whether use of optimally programmed dual-chamber ICDs lowers this risk compared with that in single-chamber ICDs is not clear.

METHODS AND

RESULTS

Subjects with a clinical indication for ICD (n=400) at 27 participating centers received dual-chamber ICDs and were randomly assigned to strictly defined optimal single- or dual-chamber detection in a single-blind manner. Programming minimized ventricular pacing. The primary end point was the proportion of SVT episodes inappropriately detected from the time of programming until crossover or end of study. On a per-episode basis, 42% of the episodes in the single-chamber arm and 69% of the episodes in the dual-chamber arm were due to SVT. Mortality (3.5% in both groups) and early study withdrawal (14% single-chamber, 11% dual-chamber) were similar in both groups. The rate of inappropriate detection of SVT was 39.5% in the single-chamber detection arm compared with 30.9% in the dual-chamber arm. The odds of inappropriate detection were decreased by almost half with the use of the dual-chamber detection enhancements (odds ratio, 0.53; 95% confidence interval, 0.30 to 0.94; P=0.03).

CONCLUSIONS

Dual-chamber ICDs, programmed to optimize detection enhancements and to minimize ventricular pacing, significantly decrease inappropriate detection.

Для цитирования: Friedman Paul A, McClelland Robyn L, Bamlet William R, Acosta Helbert, Kessler David, Munger Thomas M, Kavesh Neal G, Wood Mark, Daoud Emile, Massumi Ali, Schuger Claudio, Shorofsky Stephen, Wilkoff Bruce, Glikson Michael. Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis: the detect supraventricular tachycardia study.. Circulation. 2006. DOI: 10.1161/circulationaha.105.594531.