[Effectiveness evaluation of dynamic automatic identification system (device) of Schistosoma japonicum miracidia].

Yuan Gao, Pei-Cai Yang, Su Yang, Fei Xu

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To develop a dynamic automatic identification system (device) of Schistosoma japonicum miracidia to achieve the automatic detection and improve the detection rate and efficiency of schistosome miracidia.

METHODS

The dynamic automatic identification system (device) of S.japonicum miracidia was composed of an optical stereoscope, a digital camera, dynamic automatic tracking recognition software, and a computer. Under different conditions, the detection rates and efficiency of the system were compared with those of five professional persons.

RESULTS

The basic function of the system was to identify, label and warn the miracidia of S. japonicum, and the records could be saved automatically. The laboratory tests showed that the missing rate of the system was 0. The total consistent rates of the manual detection were 74.67% and 66.67% in the condition with and without water bug, while the total consistent rates of the system were 100.00% and 96.67%, respectively (χ2 = 9.634, 11.081, both P < 0.01) .

CONCLUSIONS

The system is much superior to manual detection in the accuracy and speed, and the system could completely replace the manual detection. Therefore, the system could be used in the field and basic research of schistosomiasis.

Для цитирования: Yuan Gao, Pei-Cai Yang, Su Yang, Fei Xu. [Effectiveness evaluation of dynamic automatic identification system (device) of Schistosoma japonicum miracidia].. Zhongguo xue xi chong bing fang zhi za zhi = Chinese journal of schistosomiasis control. 2018. DOI: 10.16250/j.32.1374.2017138.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.