DOI 10.1007/s10840-023-01618-5
Substrate and arrhythmia characterization using the multi-electrode Optrell mapping catheter for ventricular arrhythmia ablation-a single-center experience.
Аннотация
BACKGROUND
The use of a multi-electrode Optrell mapping catheter during ventricular tachycardia (VT) or premature ventricular complex (PVC) ablation procedures has not been widely reported.
OBJECTIVES
We aim to describe the feasibility and safety of using the Optrell multipolar mapping catheter (MPMC) to guide catheter ablation of VT and PVCs.
METHODS
We conducted a single-center, retrospective evaluation of patients who underwent VT or PVC ablation between June and November 2022 utilizing the MPMC.
RESULTS
A total of 20 patients met the inclusion criteria (13 VT and 7 PVC ablations, 80% male, 61 ± 15 years). High-density mapping was performed in the VT procedures with median 2753 points [IQR 1471-17,024] collected in the endocardium and 12,830 points [IQR 2319-30,010] in the epicardium. Operators noted challenges in manipulation of the MPMC in trabeculated endocardial regions or near valve apparatus. Late potentials (LPs) were detected in 11 cases, 7 of which had evidence of isochronal crowding demonstrated during late annotation mapping. Two patients who also underwent entrainment mapping had critical circuitry confirmed in regions of isochronal crowding. In the PVC group, high-density voltage and activation mapping was performed with a median 1058 points [IQR 534-3582] collected in the endocardium.
CONCLUSIONS
This novel MPMC can be used safely and effectively to create high-density maps in LV endocardium or epicardium. Limitations of the catheter include a longer wait time for matrix formation prior to starting point collection and challenges in manipulation in certain regions.