Dry suction water seal system for management of pericardial fluid during epicardial ablation.

Powell Benjamin, Coons Tricia, Lesiczka Magdalena, Markert Cate, Mehta Rohit, Misra Satish

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Epicardial ablation is an important approach in the management of patients with complex ventricular arrhythmias. Irrigated ablation catheters present a challenge in this potential space due to fluid accumulation that can cause hemodynamic compromise, requiring frequent manual fluid aspiration. In this series, we report our initial experience with the use of a dry suction water seal system for pericardial fluid management during epicardial ablation.

METHODS

Consecutive patients undergoing epicardial ventricular tachycardia (VT) ablation at a single center were included. All patients underwent epicardial access via a subxiphoid approach with a single operator. A deflectable sheath was advanced into the pericardial space, and the side port was attached to a dry suction water seal system attached to wall suction at -20 mmHg. Procedural information including patient characteristics, outcomes, and adverse events. After a period of initial experience, pericardial fluid infusion and aspiration volumes were recorded.

RESULTS

Eleven patients were included in this series. All patients underwent epicardial ablation with complete success achieved in 8 of the 11 patients and partial success in the remaining patients. Pericardial fluid intake ranging from 485 to 3050 mL with aspiration of 350-3050 mL using the dry suction water seal system. No adverse events occurred.

CONCLUSION

Dry suction water seal drainage systems can provide a safe strategy for efficient pericardial fluid management during epicardial VT ablation, potentially shortening procedure duration.

Для цитирования: Powell Benjamin, Coons Tricia, Lesiczka Magdalena, Markert Cate, Mehta Rohit, Misra Satish. Dry suction water seal system for management of pericardial fluid during epicardial ablation.. Journal of cardiovascular electrophysiology. 2023. DOI: 10.1111/jce.16017.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.