Postapproval Study of a Subcutaneous Implantable Cardioverter-Defibrillator System.

Многоцентровое исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Gold Michael R, El-Chami Mikhael F, Burke Martin C, Upadhyay Gaurav A, Niebauer Mark J, Prutkin Jordan M, Herre John M, Kutalek Steven, Dinerman Jay L, Knight Bradley P, Leigh Jill, Lucas Lindsay, Carter Nathan, Brisben Amy J, Aasbo Johan D, Weiss Raul, S-ICD System Post Approval Study Investigators

Аннотация

BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (S-ICD) was developed to avoid complications related to transvenous implantable cardioverter-defibrillator (TV-ICD) leads. Device safety and efficacy were demonstrated previously with atypical clinical patients or limited follow-up.

OBJECTIVES: The S-ICD PAS (Subcutaneous Implantable Cardioverter-Defibrillator System Post Approval Study) is a real-world, multicenter, registry of U.S. centers that was designed to assess long-term S-ICD safety and efficacy in a diverse group of patients and implantation centers.

METHODS: Patients were enrolled in 86 U.S. centers with standard S-ICD indications and were observed for up to 5 years. Efficacy endpoints were first and final shock efficacy. Safety endpoints were complications directly related to the S-ICD system or implantation procedure. Endpoints were assessed using prespecified performance goals.

RESULTS: A total of 1,643 patients were prospectively enrolled, with a median follow-up of 4.2 years. All prespecified safety and efficacy endpoint goals were met. Shock efficacy rates for discrete episodes of ventricular tachycardia or ventricular fibrillation were 98.4%, and they did not differ significantly across follow-up years (P = 0.68). S-ICD-related and electrode-related complication-free rates were 93.4% and 99.3%, respectively. Only 1.6% of patients had their devices replaced by a TV-ICD for a pacing need. Cumulative all-cause mortality was 21.7%.

CONCLUSIONS: In the largest prospective study of the S-ICD to date, all study endpoints were met, despite a cohort with more comorbidities than in most previous trials. Complication rates were low and shock efficacy was high. These results demonstrate the 5-year S-ICD safety and efficacy for a large, diverse cohort of S-ICD recipients. (Subcutaneous Implantable Cardioverter-Defibrillator [S-ICD] System Post Approval Study [PAS]; NCT01736618).