Risk of occurrence of electromagnetic interference from the application of transcutaneous electrical nerve stimulation on the sensing function of implantable defibrillators.
Аннотация
BACKGROUND
Transcutaneous electrical nerve stimulation (TENS) is an established method for pain relief. But electrical TENS currents are also a source of electromagnetic interference (EMI). Thus, TENS is considered to be contraindicated in implantable cardioverter-defibrillator (ICD) patients. However, data might be outdated due to considerable advances in ICD and cardiac resynchronization therapy (CRT) filtering and noise protection algorithm technologies. The aim of this pilot safety study was to re-evaluate the safety of TENS in patients with modern ICDs.
METHODS AND
RESULTS
One hundred and seven patients equipped with 55 different models of ICD/CRT with defibrillators from 4 manufacturers underwent a standardized test protocol including TENS at the cervical spine and the thorax, at 2 stimulation modes-high-frequency TENS (80 Hz) and burst-mode TENS (2 Hz). Potential interference monitoring included continuous documentation of ECG Lead II, intracardiac electrograms and the marker channel. Electromagnetic interference was detected in 17 of 107 patients (15.9%). Most frequent were: interpretations as a premature ventricular beats (VS/S) in 15 patients (14%), noise reversion in 5 (4.6%) which resulted in temporary asynchronous pacing in 3 (2.8%), interpretation as ventricular tachycardia/ventricular fibrillation in 2 (1.9%), and premature atrial beat in 2 (1.9%) patients. Electromagnetic interference occurrence was influenced by position (chest, P < 0.01), higher current intensity (P < 0.01), and manufacturer (P = 0.012).
CONCLUSION
Overall, only intermittent and minor EMI were detected. Prior to the use of TENS in patients with ICDs, they should undergo testing under the supervision of a cardiac device specialist.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.