Aneurysm Tears Caused by an Aneurysm Clip Springing from the Clip Applier.

Murai Yasuo, Shirokane Kazutaka, Kitamura Takao, Matano Fumihiro, Morita Akio

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

We experienced two cases in which aneurysm clips sprang from the applier. In case 1, a subdural haematoma from a ruptured anterior cerebral artery aneurysm was detected. When the clip was opened for final positioning, it suddenly sprang from the applier and ruptured the aneurysm. In case 2, the clip suddenly sprang from the applier as the surgeon opened the applier to clip an unruptured anterior cerebral aneurysm. These accidental phenomena are rare but dangerous. We present these cases to help prevent similar occurrences in the future. Video recordings of actual procedures can point to potential mechanisms and help reduce the incidence of this complication.

Для цитирования: Murai Yasuo, Shirokane Kazutaka, Kitamura Takao, Matano Fumihiro, Morita Akio. Aneurysm Tears Caused by an Aneurysm Clip Springing from the Clip Applier.. The Canadian journal of neurological sciences. Le journal canadien des sciences neurologiques. 2017. DOI: 10.1017/cjn.2016.434.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.