DOI 10.1016/s0167-5273(98)00374-x

Медицинская библиотека · International journal of cardiology · 1999

Balloon mitral valvuloplasty: comparison of haemodynamic and echocardiographic assessment of mitral stenosis at different heart rates in the catheterisation laboratory.

Harcombe A A, Ludman P F, Wisbey C, Crowley J J, Sharples L, Shapiro L M

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

AIMS: To compare echo-Doppler, Gorlin equation and haemodynamic methods of measuring mitral valve stenosis during right ventricular pacing-induced tachycardia before and after Inoue balloon mitral valvuloplasty to determine which method gave the most consistent results.

METHODS AND

RESULTS

Measurements were made before and after valvuloplasty at: baseline heart rates, paced at 115 and then 145 beats/min. Mitral valve area by echo-Doppler was 1.1(+/-0.1) cm2 (mean +/- S.E.) before and 1.8(+/-0.2) cm2 after valvuloplasty; and by Gorlin equation: 0.9(+/-0.1) cm2 before and 1.5(+/-0.1) cm2 after. Echo-Doppler measurements were heart rate dependent but those by Gorlin measurements were not. At baseline, cardiac index was 2.08(+/-0.2) l min(-1), left atrial pressure 23.3(+/-7.9) mm Hg and mean mitral diastolic gradient 16.9(+/-9.9) mm Hg. After valvuloplasty, cardiac index was 2.31(+/-0.1) l min(-1), left atrial pressure fell to 19.2(+/-5.6) mm Hg and mean diastolic gradient was reduced to 8.5(+/-1.8) mm Hg.

CONCLUSIONS

The Gorlin mitral valve area appeared to be the most heart rate independent indicator of success following valvuloplasty.

Для цитирования: Harcombe A A, Ludman P F, Wisbey C, Crowley J J, Sharples L, Shapiro L M. Balloon mitral valvuloplasty: comparison of haemodynamic and echocardiographic assessment of mitral stenosis at different heart rates in the catheterisation laboratory.. International journal of cardiology. 1999. DOI: 10.1016/s0167-5273(98)00374-x.