Managing Catheter Knotting in Pulse-Select Pulsed Field Ablation: Mechanisms and a Practical Strategy for Resolution.
Аннотация
BACKGROUND
Pulsed field ablation (PFA) using the PulseSelect catheter is effective for atrial fibrillation but may occasionally result in catheter deformation or knotting.
METHODS
We analyzed 25 PulseSelect catheters used for AF ablation to identify knotting mechanisms. Two reproducible knot types were observed. A resolution strategy was tested: retraction to the sheath tip, wire withdrawal, and gradual advancement with or without rotation.
RESULTS
The knot was resolved in 24 catheters (96%), with 21 (84%) resolved on the first attempt. One severe type 1 knot was resistant but safely removed.
CONCLUSIONS
A structured, non-forceful approach utilizing the catheter's elasticity effectively resolves knotting without rendering the catheter unusable.