MedElevator Войти в исследования
← Поиск публикаций

DOI 10.4244/eij-d-20-00763

Медицинская библиотека · EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology · 2020

Twelve-month clinical and imaging outcomes of the uncaging coronary DynamX bioadaptor system.

Verheye Stefan, Vrolix Mathias, Montorfano Matteo, Zivelonghi Carlo, Giannini Francesco, Bedogni Francesco, Dubois Christophe, De Bruyne Bernard, Costa Ricardo A, Chamié Daniel, de Ribamar Costa José, Kereiakes Dean J, Abizaid Alexandre A, Colombo Antonio

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

AIMS: We aimed to assess the safety and efficacy of the DynamX Novolimus-Eluting Coronary Bioadaptor System, a novel device that initially acts as a second-generation drug-eluting stent, but after six months frees the vessel through uncaging elements.

METHODS AND

RESULTS

This multicentre study enrolled 50 patients with single de novo lesions. In-device acute lumen gain was 1.61±0.34 mm, and device and procedure success was 100%. Up to 12 months, two target lesion failures occurred: both were cardiac deaths (day 255 and day 267 post procedure). No definite or probable device thrombosis was observed. Mean late lumen loss was 0.12±0.18 mm in-device and 0.11±0.16 mm in-segment. Per intravascular ultrasound, the mean device area and mean vessel area increased significantly by 5% and 3%, respectively, while the mean lumen area was maintained. Stationary optical coherence tomography in seven patients demonstrated restoration of cyclic pulsatility, with an approximate lumen area variance of 11% between systole and diastole.

CONCLUSIONS

The DynamX bioadaptor showed drug-eluting stent-like acute performance and safety and efficacy up to one year. Positive remodelling with an increase of vessel and device area while maintaining the mean lumen area was demonstrated. Long-term follow-up and randomised trials are required to assess the benefit of this device on events beyond one year.

Для цитирования: Verheye Stefan, Vrolix Mathias, Montorfano Matteo, Zivelonghi Carlo, Giannini Francesco, Bedogni Francesco, Dubois Christophe, De Bruyne Bernard, Costa Ricardo A, Chamié Daniel, de Ribamar Costa José, Kereiakes Dean J, Abizaid Alexandre A, Colombo Antonio. Twelve-month clinical and imaging outcomes of the uncaging coronary DynamX bioadaptor system.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2020. DOI: 10.4244/eij-d-20-00763.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.