Медицинская библиотека · International journal of sports physical therapy · 2023

Reliability and Concurrent Validity of a Markerless, Single Camera, Portable 3D Motion Capture System for Assessment of Glenohumeral Mobility.

Pottorf Ofra, Vapne Daniel, Ghigiarelli Jamie, Haase Kaitlyn

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Recent technological advancements have enabled medical, sport, and fitness professionals to utilize digital tools that assist with conducting movement examinations and screenings. One such advancement has been the implementation of a single camera, markerless, and portable 3D motion capture system designed to obtain ROM measurements for multiple body parts simultaneously. However, the reliability and validity of a markerless 3D motion capture system that uses a single camera has not been established.

PURPOSE: The purpose of this study was to investigate the reliability and concurrent validity of this 3D motion capture system compared to a goniometer in assessing ROM of the glenohumeral joint.

STUDY

DESIGN

Quasi-experimental reliability, convenience sampling.

METHODS/MATERIALS: Forty healthy volunteers (mean ± SD, age 27.4 ± 12.4 years, height 173.4 ± 11.1 cm, weight 72.1 ± 16.2 kg) participated in this study. Intrarater reliability was analyzed by ICC(2,k) with a 95% CI using two repeated trials for each shoulder movement (flexion, abduction, external rotation, internal rotation) that were recorded simultaneously via two methods: a standard goniometer and a 3D motion capture system. Concurrent validity was analyzed using Pearson Correlation Coefficient (r).

RESULTS

The intrarater reliability between the two instruments for glenohumeral motions yielded an overall ICC of 0.82 (95% CI: 0.74-0.88) indicating good reliability for both instruments.The 3D motion capture system demonstrated strong correlations with goniometry for shoulder flexion (r = 0.67), abduction (r = 0.63), and external rotation (r = 0.76), and very strong correlation for shoulder internal rotation (r = 0.84).

CONCLUSION

Results from this study indicated that a markerless, single camera, portable 3D motion capture system can be a reliable and valid tool to assess glenohumeral joint ROM in comparison to a standard goniometer.

LEVEL OF EVIDENCE: 3.

Для цитирования: Pottorf Ofra, Vapne Daniel, Ghigiarelli Jamie, Haase Kaitlyn. Reliability and Concurrent Validity of a Markerless, Single Camera, Portable 3D Motion Capture System for Assessment of Glenohumeral Mobility.. International journal of sports physical therapy. 2023. DOI: 10.26603/001c.88003.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.