Performance of the Dexcom G6 Continuous Glucose Monitoring System During Cardiac Surgery Using Hypothermic Extracorporeal Circulation.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Herzig David, Vettoretti Martina, Guensch Dominik P, Melmer Andreas, Schürch Daniel, Roos Jonathan, Goerg Arna M C, Krutkyte Gabija, Cecchini Luca, Facchinetti Andrea, Vogt Andreas P, Bally Lia
Аннотация
OBJECTIVE: Continuous glucose monitoring (CGM) may be challenged by extreme conditions during cardiac surgery using hypothermic extracorporeal circulation (ECC).
RESEARCH DESIGN AND METHODS: We evaluated the Dexcom G6 sensor in 16 subjects undergoing cardiac surgery with hypothermic ECC, of whom 11 received deep hypothermic circulatory arrest (DHCA). Arterial blood glucose, quantified by the Accu-Chek Inform II meter, served as reference.
RESULTS: Intrasurgery mean absolute relative difference (MARD) of 256 paired CGM/reference values was 23.8%. MARD was 29.1% during ECC (154 pairs) and 41.6% immediately after DHCA (10 pairs), with a negative bias (signed relative difference: -13.7%, -26.6%, and -41.6%). During surgery, 86.3% pairs were in Clarke error grid zones A or B and 41.0% of sensor readings fulfilled the International Organization for Standardization (ISO) 15197:2013 norm. Postsurgery, MARD was 15.0%.
CONCLUSIONS: Cardiac surgery using hypothermic ECC challenges the accuracy of the Dexcom G6 CGM although recovery appears to occur thereafter.
RESEARCH DESIGN AND METHODS: We evaluated the Dexcom G6 sensor in 16 subjects undergoing cardiac surgery with hypothermic ECC, of whom 11 received deep hypothermic circulatory arrest (DHCA). Arterial blood glucose, quantified by the Accu-Chek Inform II meter, served as reference.
RESULTS: Intrasurgery mean absolute relative difference (MARD) of 256 paired CGM/reference values was 23.8%. MARD was 29.1% during ECC (154 pairs) and 41.6% immediately after DHCA (10 pairs), with a negative bias (signed relative difference: -13.7%, -26.6%, and -41.6%). During surgery, 86.3% pairs were in Clarke error grid zones A or B and 41.0% of sensor readings fulfilled the International Organization for Standardization (ISO) 15197:2013 norm. Postsurgery, MARD was 15.0%.
CONCLUSIONS: Cardiac surgery using hypothermic ECC challenges the accuracy of the Dexcom G6 CGM although recovery appears to occur thereafter.