Journal of clinical monitoring and computing · 2015
Agreement between stroke volume measured by oesophageal Doppler and uncalibrated pulse contour analysis during fluid loads in severe aortic stenosis.
СтатьяНаблюдательное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Høiseth Lars Øivind, Hoff Ingrid Elise, Hagen Ove Andreas, Landsverk Svein Aslak, Kirkebøen Knut Arvid
Аннотация
The purpose of this analysis was to study agreement and trending of stroke volume measured by oesophageal Doppler and 3rd generation Vigileo during fluid loads in patients with severe aortic stenosis. Observational study in 32 patients (30 analyzed) scheduled for aortic valve replacement due to severe aortic stenosis. After induction of anesthesia and before start of surgery, hemodynamic registrations for 1 min were obtained before and after a fluid load. Agreement between stroke volume measured by oesophageal Doppler (SVOD) and Vigileo (SVVig) was evaluated in Bland-Altman plot and trending in four-quadrant and polar plots. Bias ± limits of agreement (LOA) between SVOD and SVVig was 24 ± 37 ml (percentage error 45%). Concordance of the two methods from before to after a fluid load was 100%. Angular bias ± LOA was 12° ± 28°. Absolute values of SVOD and SVVig agreed poorly, but changes were highly concordant during fluid loads in aortic stenosis patients. The angular agreement indicated acceptable trending. The two measurement methods are not interchangeable in patients with aortic stenosis.
Предметные рубрики MeSH
AgedAlgorithmsAnestheticsAortic ValveAortic Valve StenosisCardiac OutputCardiovascular DiseasesCoronary Artery DiseaseDiabetes MellitusEsophagusFemaleHeart RateHeart Valve ProsthesisHemodynamicsHumansHypertensionMaleMiddle AgedMonitoring, IntraoperativePressureReproducibility of ResultsSignal Processing, Computer-AssistedStroke VolumeUltrasonography, Doppler