Identification of "substrate fibrillators" and "trigger fibrillators" by pacemaker diagnostics.

Yang Alexander, Ruiter Jaap, Pfeiffer Dietrich, Esmailzadeh Bahman, Welz Armin, Hochhäusler Marc, Nickenig Georg, Lüderitz Berndt, Lewalter Thorsten

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The population of patients likely to respond to selected pacing algorithms for maintenance of sinus rhythm is unknown.

OBJECTIVES

The purpose of this study was to identify patients with specific onset patterns of paroxysmal atrial fibrillation (AF).

METHODS

Dual-chamber pacemakers with advanced diagnostic functions were implanted in 112 patients with conventional indications for antibradycardia pacing and a history of paroxysmal AF. Pacemaker diagnostic data were analyzed after 97.5 +/- 40.9 days. According to the frequency of premature atrial contractions (PACs) during the 5 minutes before AF onset, patients were assigned to one of three groups: group A (high PAC activity), group B (moderate PAC activity), or group C (low PAC activity).

RESULTS

AF burden was lower in group A (4.6% +/- 2.4%) than group B (15.8% +/- 3.0%, P = .003) and group C (15.5% +/- 3.1%, P = .003). Fewer AF episodes occurred in group A (2.1 +/- 1.3 per day) than group B (3.8 +/- 1.2 per day, P = .006). Mean AF episode duration was shorter in group A (11.4 +/- 10.2 hours) than group C (41.4 +/- 27.5 hours, P = .03).

CONCLUSION

The coincidence of low PAC activity before AF onset, high AF burden, and extended arrhythmia episode duration appears to be the consequence of a high atrial substrate factor. In these "substrate fibrillators," the efficacy of pacing algorithms for maintenance of sinus rhythm may be limited. In contrast, "trigger fibrillators" exhibiting low AF burden despite high PAC incidence may represent the target population for specific PAC-suppressing pacing algorithms.

Для цитирования: Yang Alexander, Ruiter Jaap, Pfeiffer Dietrich, Esmailzadeh Bahman, Welz Armin, Hochhäusler Marc, Nickenig Georg, Lüderitz Berndt, Lewalter Thorsten. Identification of "substrate fibrillators" and "trigger fibrillators" by pacemaker diagnostics.. Heart rhythm. 2006. DOI: 10.1016/j.hrthm.2006.01.030.