MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/hjh.0000000000003632

Медицинская библиотека · Journal of hypertension · 2023

Simultaneous 24-h ambulatory blood pressure measurement on both arms: a consideration for improving hypertension management.

Bothe Tomas L, Kreutz Reinhold, Glos Martin, Patzak Andreas, Pilz Niklas

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Arterial hypertension is one of the common treatment goals in today's medicine. 24-h ambulatory blood pressure measurement (ABPM) performed by oscillometric cuff-based devices is considered as the gold standard in hypertension diagnostics. This study aims at examining the measurement accuracy of a widely used, ABPM device.

METHODS

Fifty-two young and healthy participants underwent simultaneous 24-h ABPM on the left and the right upper arm using two Boso/A&D TM-2430 oscillometric cuff-based devices. Pressure curves of the cuffs, as well as hydrostatic pressure difference between the cuffs were recorded.

RESULTS

The mean differences between both simultaneous measurements were 1.16 mmHg with limits of agreement of 36.23 mmHg for SBP and 1.32 mmHg with limits of agreement of 32.65 mmHg for DBP. Excluding measurements where the pressure curves were disturbed and correcting for hydrostatic pressure difference between the cuffs, reduced the measurement error. However, limits of agreement remained around 20 mmHg. There were large differences in hypertension grading and dipping pattern classification between simultaneous measurements on the left and right arm.

CONCLUSION

The cuff-based ABPM device reveals notable measurement uncertainties, influencing hypertension grading, dipping pattern classification and blood pressure variability. These effects are attributed in part to disturbances during cuff deflation and hydrostatic influences. Nonetheless, ABPM has shown its clinical values in several studies, while this study underscores its still unlocked potential to improve hypertension management.

Для цитирования: Bothe Tomas L, Kreutz Reinhold, Glos Martin, Patzak Andreas, Pilz Niklas. Simultaneous 24-h ambulatory blood pressure measurement on both arms: a consideration for improving hypertension management.. Journal of hypertension. 2023. DOI: 10.1097/hjh.0000000000003632.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.