Progressive Device Failure at Long Term Follow Up of the Nellix EndoVascular Aneurysm Sealing (EVAS) System.

Singh Aminder A, Benaragama Kapila S, Pope Tasneem, Coughlin Patrick A, Winterbottom Andrew P, Harrison Seamus C, Boyle Jonathan R

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

High rates of midterm failure of the Nellix EndoVascular Aneurysm Sealing (EVAS) System resulted in device withdrawal from the UK market. The study aim was to report long term Nellix EVAS outcomes and management of a failing device.

METHODS

A retrospective review of EVAS procedures at a tertiary unit was performed. Device failure was defined as a triad of stent migration, stent separation, and secondary sac expansion, or any intervention for type 1 endoleak, device rupture, or explant.

RESULTS

161 (male n = 140, female n = 21) patients with a median follow up of 6.0 (IQR 5.0-6.6) years were included. Freedom from all cause mortality estimate at six years was 41.5%. There were 70 (43.5%) device failures with a freedom from device failure estimate at six years of 32.3%. Failure was the result of sac expansion (n = 41), caudal stent migration (n = 36), stent separation (n = 26), and secondary AAA rupture (n = 15). A substantial number of type 1 endoleaks was present (1a n = 33, 1b n = 11), but the type 2 endoleak rate was low at 3.7%. Some 36 (22.4%) patients required re-intervention. Twenty-one patients underwent explant with no 30 day deaths. Six patients underwent Nellix-in-Nellix application (NINA) with one early death from bowel ischaemia and one patient who died later from non-aneurysm related cause. Two NINA patients have ongoing sac expansion and two have had thrombosis of a Nellix limb or visceral stent. Proximal embolisation was only successful in one of six cases.

CONCLUSION

The long term failure rate of Nellix EVAS is high. All patients with a device must be informed and be enrolled in enhanced surveillance. EVAS explant is an acceptable technique with favourable outcomes. Management by open explant, if the patient is fit, should be considered early and offered to those with device failure.

Для цитирования: Singh Aminder A, Benaragama Kapila S, Pope Tasneem, Coughlin Patrick A, Winterbottom Andrew P, Harrison Seamus C, Boyle Jonathan R. Progressive Device Failure at Long Term Follow Up of the Nellix EndoVascular Aneurysm Sealing (EVAS) System.. European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery. 2020. DOI: 10.1016/j.ejvs.2020.11.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.