Novel Colorimetric and Light Scatter Methods to Identify and Manage Peritoneal Dialysis-Associated Peritonitis at the Point-of-Care.

Govindji-Bhatt Nishal, Kennedy Stephnie M, Barker Michael G, Kell Darren, Henderson Duncan, Goddard Nicholas, Garcia Ana Yepes, Milner Adam S, Willett Tom, Griffiths Ryan, Foster Peter, Kilgallon William, Cant Rachel, Knight Christopher G, Lewis David, Corbett Richard, Akbani Habib, Woodrow Graham, Sood Bhrigu, Iyasere Osasuyi, Davies Simon, Qazi Junaid, Vardhan Anand, Gillis Laura, Wilkie Martin, Dobson Curtis B

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Peritoneal dialysis (PD)-related peritonitis (PDRP) is a common cause of transfer to hemodialysis, patient morbidity, and is a risk factor for mortality. Associated patient anxiety can deter selection of PD for renal replacement therapy. Diagnosis relies on hospital laboratory tests; however, this might be achieved earlier if such information was available at the point-of-care (POC), thereby significantly improving outcomes. The presence of culturable microbes and the concentration of leukocytes in effluent both aid peritonitis diagnosis, as specified in the International Society for Peritoneal Dialysis (ISPD) diagnostic guidelines. Here, we report the development of 2 new methods providing such information in simple POC tests.

METHODS

One approach uses a tetrazolium-based chemical reporting system, primarily focused on detecting bacterial contamination and associated vancomycin-sensitivity. The second approach uses a novel forward light-scatter device (QuickCheck) to provide an instant quantitative cell count directly from PD patient effluent.

RESULTS

The tetrazolium approach detected and correctly distinguished laboratory isolates, taking 10 hours to provide non-quantitative results. We compared the technical performance of the light scatter leukocyte counting approach with spectrophotometry, hemocytometer counting and flow cytometry (Sysmex) using patient effluent samples. QuickCheck had high accuracy (94%) and was the most precise (coefficient of variation <4%), showing minimal bias, overall performing similarly to flow cytometry.

CONCLUSION

These complementary new approaches provide a simple means to obtain information to assist diagnosis at the POC. The first provides antibiotic sensitivity following 10 hours incubation, whereas the second optical approach (QuickCheck), provides instant accurate total leukocyte count.

Для цитирования: Govindji-Bhatt Nishal, Kennedy Stephnie M, Barker Michael G, Kell Darren, Henderson Duncan, Goddard Nicholas, Garcia Ana Yepes, Milner Adam S, Willett Tom, Griffiths Ryan, Foster Peter, Kilgallon William, Cant Rachel, Knight Christopher G, Lewis David, Corbett Richard, Akbani Habib, Woodrow Graham, Sood Bhrigu, Iyasere Osasuyi, Davies Simon, Qazi Junaid, Vardhan Anand, Gillis Laura, Wilkie Martin, Dobson Curtis B. Novel Colorimetric and Light Scatter Methods to Identify and Manage Peritoneal Dialysis-Associated Peritonitis at the Point-of-Care.. Kidney international reports. 2023. DOI: 10.1016/j.ekir.2023.12.021.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.