MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jvs.2015.08.093

Медицинская библиотека · Journal of vascular surgery · 2015

One-year outcomes of the U.S. and Japanese regulatory trial of the Misago stent for treatment of superficial femoral artery disease (OSPREY study).

Ohki Takao, Angle John F, Yokoi Hiroyoshi, Jaff Michael R, Popma Jeffrey, Piegari Guy, Kanaoka Yuji, OSPREY investigators

Клиническое исследованиеСравнительное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

The purpose of this study was to assess the safety and efficacy of the Misago stent (Terumo Corp, Tokyo, Japan) in occlusive and stenotic superficial femoral artery (SFA) disease.

METHODS

The safety and efficacy of the Misago SFA stent were evaluated prospectively in this initial collaboration trial between Japan and the United States. Because this trial enrolled patients mainly from Japan and the United States and because there is a question as to whether a race difference exists in SFA stent performance, the race difference on outcome was also analyzed. In addition, results were compared with a prior SFA stent trial.

RESULTS

The Misago stent was implanted in 261 subjects with TransAtlantic Inter-Society Consensus (TASC) type A and type B SFA lesions (201 subjects in the United States, 50 in Japan, 9 in Taiwan, 1 in South Korea). The mean age of the patients was 69.3 ± 10.0 years, and the mean lesion length was 83.8 ± 41.3 mm. The overall 12-month primary patency rate and clinically driven target lesion revascularization were 82.9% and 13.0%, respectively. Regional differences within the Occlusive/Stenotic Peripheral Artery Revascularization Study (OSPREY) and outcomes between U.S. and Asian patients were similar, including primary patency (82.9% vs 83.0%; P = .889), clinically driven target lesion revascularization (13.4% vs 11.7%; P = .829), stent fracture rate (1.3% vs 0.0%; P = 1.000), and stent thrombosis rate (0.5% vs 0.0%; P = 1.000).

CONCLUSIONS

OSPREY 12-month data showed satisfactory outcome of the Misago stent for the treatment of TASC type A and type B SFA lesions and appears to be comparable to recent stent trials. In addition, the lack of difference in outcome among races supports the value of international trials.

Для цитирования: Ohki Takao, Angle John F, Yokoi Hiroyoshi, Jaff Michael R, Popma Jeffrey, Piegari Guy, Kanaoka Yuji, OSPREY investigators. One-year outcomes of the U.S. and Japanese regulatory trial of the Misago stent for treatment of superficial femoral artery disease (OSPREY study).. Journal of vascular surgery. 2015. DOI: 10.1016/j.jvs.2015.08.093.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.