DOI 10.3109/0886022x.2015.1077316
Doppler ultrasound and calcification score: improving vascular access surveillance.
Аннотация
AIM: Vascular access (VA) dysfunction limits hemodialysis delivery, which increases morbidity and mortality. The most com mon cause of VA failure is thrombosis, due to flow limiting stenosis resulting from neointimal hyperplasia. This occurs not only due to hemodynamic factors but also by systemic ones related to vascular atherosclerosis, inflammation and calcification, which has developed a simple vascular calcification score (SVCS) predictor of vascular calcification and arterial stiffness. The NKF-K/DOQ recommends several diagnostic procedures for VA surveillance. Blood access flow (Qa) has predictive power for the detection of stenosis. Our aim was to evaluate the role of systemic factors, especially SCVS, on Qa.
MATERIAL AND
METHODS
Transversal study in 50 patients. Qa value was obtained with Blood Temperature Monitor and Doppler method. Pearson coefficient evaluated correlation between them. Clinical, lab and radiological variables were recorded and non-parametric tests evaluated how both Qa varied with them.
RESULTS
Pearson's corelation between DU-Qa and TD-Qa was 0.851 (p-value <0.001). DU-Qa varied significantly with age (p = 0.012), VA type (p = 0.021), SCVS (p = 0.030), intra-access arterial pressure (p = 0.015) and time on dialysis (p = 0.002). BTM-Qa varied significantly with diabetes status (p = 0.027), age (p = 0.017), first VA status (p = 0.036), intra-access arterial pressure (p = 0.028) and dialysis time (p = 0.001). Nevertheless, gender, hypertensive status and analitical parameters did not change the flow values.
CONCLUSION
Higher SVCS was associated only with lower DU-Qas, giving this method an advantage towards the indirect one. Additionally, a simple method like SVCS may be used to guide new surveillance recommendations accordingly to risk stratification.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.