DOI 10.1016/j.jcin.2015.03.020

Медицинская библиотека · JACC. Cardiovascular interventions · 2015

Impact of Paclitaxel Dose on Tissue Pharmacokinetics and Vascular Healing: A Comparative Drug-Coated Balloon Study in the Familial Hypercholesterolemic Swine Model of Superficial Femoral In-Stent Restenosis.

Gongora Carlos A, Shibuya Masahiko, Wessler Jeffrey D, McGregor Jenn, Tellez Armando, Cheng Yanping, Conditt Gerard B, Kaluza Greg L, Granada Juan F

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study sought to compare the effect of paclitaxel-coated balloon (PCB) concentration on tissue levels and vascular healing using 3 different PCB technologies (In.Pact Pacific = 3 μg/mm(2), Lutonix = 2 μg/mm(2) and Ranger = 2 μg/mm(2)) in the experimental setting.

BACKGROUND

The optimal therapeutic dose for PCB use has not been determined yet.

METHODS

Paclitaxel tissue levels were measured up to 60 days following PCB inflation (Ranger and In.Pact Pacific) in the superficial femoral artery of healthy swine (18 swine, 36 vessels). The familial hypercholesterolemic swine model of superficial femoral artery in-stent restenosis (6 swine, 24 vessels) was used in the efficacy study. Two weeks following bare-metal stent implantation, each in-stent restenosis site was randomly treated with a PCB or an uncoated control balloon (Sterling). Quantitative vascular analysis and histology evaluation was performed 28 days following PCB treatment.

RESULTS

All PCB technologies displayed comparable paclitaxel tissue levels 4 h following balloon inflation. At 28 days, all PCB had achieved therapeutic tissue levels; however, the In.Pact PCB resulted in higher tissue concentrations than did the other PCB groups at all time points. Neointimal inhibition by histology was decreased in all PCB groups compared with the control group, with a greater decrease in the In.Pact group. However, the neointima was more mature and contained less peri-strut fibrin deposits in both 2-μg/mm(2) PCB groups.

CONCLUSIONS

Compared with the clinically established PCB dose, lower-dose PCB technologies achieve lower long-term tissue levels but comparable degrees of neointimal inhibition and fewer fibrin deposits. The impact of these findings in restenosis reduction and clinical outcomes needs to be further investigated.

Для цитирования: Gongora Carlos A, Shibuya Masahiko, Wessler Jeffrey D, McGregor Jenn, Tellez Armando, Cheng Yanping, Conditt Gerard B, Kaluza Greg L, Granada Juan F. Impact of Paclitaxel Dose on Tissue Pharmacokinetics and Vascular Healing: A Comparative Drug-Coated Balloon Study in the Familial Hypercholesterolemic Swine Model of Superficial Femoral In-Stent Restenosis.. JACC. Cardiovascular interventions. 2015. DOI: 10.1016/j.jcin.2015.03.020.