DOI 10.1034/j.1600-0501.1995.060103.x
A histomorphometric and removal torque study of screw-shaped titanium implants with three different surface topographies.
Аннотация
Screw-shaped implants with 3 different surface topographies, evidenced visually as well as numerically with an optical profilometer, were inserted in rabbit bone. After a healing period of 12 weeks, a statistically significant higher removal torque was needed to unscrew screws blasted with 25-microns TiO2 particles and screws blasted with 75-microns particles of AI2O3 compared with screws with a turned surface. The histomorphometric evaluation demonstrated a higher percentage of bone-to-metal contact for implants blasted with 25-microns particles of TiO2 compared with the as-machined implants. A greater surface area of bone in threads was found for the turned screws compared with screws blasted with 25-microns TiO2 particles. In the short-term follow-up, there was a better fixation of implants with an average surface roughness of 0.9-1.3-microns and with a homogeneous surface structure than of implants with an average surface roughness of 0.4 microns and with a clear direction of the surface pattern.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.