Impact of heart rate on coronary artery stenosis grading accuracy using deep learning-based fast kV-switching CT: A phantom study.

Mikayama Ryoji, Kojima Tsukasa, Shirasaka Takashi, Yamane Shiho, Funatsu Ryohei, Kato Toyoyuki, Yabuuchi Hidetake

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Deep learning-based fast kV-switching CT (DL-FKSCT) generates complete sinograms for fast kV-switching dual-energy CT (DECT) scans by using a trained neural network to restore missing views. Such restoration significantly enhances the image quality of coronary CT angiography (CCTA), and the allowable heart rate (HR) may vary between DECT and single-energy CT (SECT). This study aimed to examine HR's effect onCCTA using DL-FKSCT.

METHODS

We scanned stenotic coronary artery phantoms attached to a pulsating cardiac phantom with DECT and SECT modes on a DL-FKSCT scanner. The phantom unit was operated with simulated HRs ranging from 0 (static) to 50-70 beats per minute (bpm). The sharpness and stenosis ratio of the coronary model were quantitatively compared between DECT and SECT, stratified by simulated HR settings using the paired t-test (significance was set at p < 0.01 with a Bonferroni adjustment for multiple comparisons).

RESULTS

Regarding image sharpness, DECT showed significant superiority over SECT. In terms of the stenosis ratio compared to a static image reference, 70 keV virtual monochromatic image in DECT exhibited errors exceeding 10 % at HRs surpassing 65 bpm (p < 0.01), whereas 120 kVp SECT registered errors below 10 % across all HR settings, with no significant differences observed.

CONCLUSIONS

In DL-FKSCT, DECT exhibited a lower upper limit of HR than SECT. Therefore, HR control is important for DECT scans in DL-FKSCT.

Для цитирования: Mikayama Ryoji, Kojima Tsukasa, Shirasaka Takashi, Yamane Shiho, Funatsu Ryohei, Kato Toyoyuki, Yabuuchi Hidetake. Impact of heart rate on coronary artery stenosis grading accuracy using deep learning-based fast kV-switching CT: A phantom study.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2025. DOI: 10.1016/j.ejmp.2025.105046.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.