No Difference in Pullout Strength Between a Bio-inductive Implant and a Semitendinosus Tendon Graft in a Biomechanical Study of Medial Patellofemoral Ligament Repair Augmentation.

Wetzler Austin, McMillan Sean, Brewer Erik, Patel Aakash, Handy Samuel, Wetzler Merrick

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the pullout strength of a bio-inductive implant (BI) used to augment a medial patellofemoral ligament (MPFL) repair with the pullout strength of semitendinosus graft in a biomechanical cadaveric model.

METHODS

Six matched pairs of cadavers (12 knees) were used in the biomechanical testing comparing semitendinosus tendon (Semi-T) versus a BI. The Semi-T was harvested from 1 of the matched pairs. A standard double-bundle technique using 2 sockets in the upper two-thirds of the patella 15 mm apart was performed. After docking of the graft into the patella, the patella was dissected free of soft tissues and potted into a fixture to allow mechanical pull parallel to the transverse axis of the patella. The construct was pulled to failure.

RESULTS

There was no statistically significant difference in pullout strength (P = .77) between the BI group (249.3 ± 36.3 N) and Semi-T group (235.0 ± 113.6 N) double-bundle constructs. In the Semi-T group, 50% of the specimens (3 of 6 knees) failed via anchor pullout and a fourth specimen failed at the suture-anchor interface (16.7%), whereas in the BI group, 16.7% of the specimens (1 of 6 knees) failed by anchor pullout. Although the Semi-T group (49.5 ± 14.1 N/mm) showed significantly greater stiffness than the BI group (13.8 ± 0.6 N/mm, P < .01), pullout strength in the Semi-T group was highly variable: 50% of the specimens (3 of 6 knees) with semitendinosus constructs failed at 5 mm of displacement or less via graft or anchor pullout. Maximum load, displacement at failure, stiffness, and load at 5 mm were compared between the augmented and non-augmented control specimens using a 2-tailed non-equal variance Student t test. For all comparisons, P < .05 was considered to indicate a statistically significant difference.

CONCLUSIONS

In this biomechanical study, augmentation of an MPFL reconstruction using a common double-bundle technique with a BI had the same pullout strength as a semitendinosus graft using the same technique in cadaveric knees.

CLINICAL RELEVANCE: MPFL repair after a patellar dislocation may be inadequate to restore the strength of the native MPFL and prevent recurrent patellar instability. Recurrent instability of the patella can result in progressive injury to the soft tissue and articular cartilage of the patella and femur. It is important to study the techniques used for MPFL repair to continually improve patient outcomes. Further testing of these additional techniques and clinical studies are needed to evaluate the implants used to augment MPFL repairs.

Для цитирования: Wetzler Austin, McMillan Sean, Brewer Erik, Patel Aakash, Handy Samuel, Wetzler Merrick. No Difference in Pullout Strength Between a Bio-inductive Implant and a Semitendinosus Tendon Graft in a Biomechanical Study of Medial Patellofemoral Ligament Repair Augmentation.. Arthroscopy, sports medicine, and rehabilitation. 2024. DOI: 10.1016/j.asmr.2023.100827.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.