Predictive factors of lead failure in patients implanted with cardiac devices.

Aizawa Yoshiyasu, Negishi Masachika, Kashimura Shin, Nakajima Kazuaki, Kunitomi Akira, Katsumata Yoshinori, Nishiyama Takahiko, Kimura Takehiro, Nishiyama Nobuhiro, Fukumoto Kotaro, Tanimoto Yoko, Kohsaka Shun, Takatsuki Seiji, Fukuda Keiichi

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

Lead failures (LFs) are one of the most common complications in patients implanted with cardiovascular implantable electronic devices. LFs often cause serious secondary complications such as inappropriate ICD shocks or asystole. This study aimed to identify the clinical factors associated with the occurrence of LFs.

METHODS

A total of 735 consecutive device implantations (mean age 67±15years, males 64%) performed at a single university hospital setting from 1997 to 2014 were included. The implanted devices consisted of 421 pacemakers, 250 implantable cardioverter defibrillators (ICD), 9 cardiac resynchronization therapy pacemakers (CRT-P), and 55 CRT defibrillators (CRT-D). The primary endpoint was the development of an LF.

RESULTS

During a mean duration of 5.8±4.3years, 38 LFs developed in 31 patients (mean age 56±14years). LFs included 32 ICD (7 Sprint Fidelis, 2 Riata), and 6 pacing leads. Nine patients received inappropriate ICD shocks and 1 had syncope due to an LF. All patients underwent lead reinsertions with device replacements. Eight patients required opposite site implantations due to venous occlusions. The predictive factors of LFs were the age, male sex, taller body length, ICD vs. pacemaker, lesser lead number, extra-thoracic puncture of the axillary vein vs. a cut-down of the cephalic vein, use of recalled leads and patients with idiopathic ventricular fibrillation (IVF) and Brugada syndrome (BrS).

CONCLUSION

LFs occurred mainly with ICD leads. A lesser age, the puncture method, lead model, and diagnosis of IVF/BrS were associated with the development of LFs.

Для цитирования: Aizawa Yoshiyasu, Negishi Masachika, Kashimura Shin, Nakajima Kazuaki, Kunitomi Akira, Katsumata Yoshinori, Nishiyama Takahiko, Kimura Takehiro, Nishiyama Nobuhiro, Fukumoto Kotaro, Tanimoto Yoko, Kohsaka Shun, Takatsuki Seiji, Fukuda Keiichi. Predictive factors of lead failure in patients implanted with cardiac devices.. International journal of cardiology. 2015. DOI: 10.1016/j.ijcard.2015.07.055.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.